Medtronic Neuromodulation
Class II Ongoing
Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B...
Reason: Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from ear...
Z-2201-2026 May 20, 2026 Minneapolis, MN
Dexcom, Inc.
Class II Ongoing
Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.5.0 and 1.6.0 Product Description: Dex...
Reason: A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app...
Z-2165-2026 May 20, 2026 San Diego, CA
DJO, LLC
Class II Ongoing
Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: No Product Description: 11-1424 - ICEMAN W/,UNIV,LOOP,NS,EH The IceMan CLASSIC3¿ cold t...
Reason: Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
Z-2136-2026 May 20, 2026 Carlsbad, CA
DJO, LLC
Class II Ongoing
Brand Name: DONJOY Product Name: ICEMAN CLASSIC CUBE Model/Catalog Number: 11-0494 Software Version: no Product Description: 11-0494 - ICEMAN CLASSIC CUBE The IceMan CLASSIC3¿ cold therapy unit ...
Reason: Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
Z-2134-2026 May 20, 2026 Carlsbad, CA
DJO, LLC
Class II Ongoing
Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,RH Model/Catalog Number: 11-1422 Software Version: no Product Description: 11-1422 - ICEMAN W/,UNIV,LOOP,NS,RH The IceMan CLASSIC3¿ cold t...
Reason: Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
Z-2135-2026 May 20, 2026 Carlsbad, CA
Dexcom, Inc.
Class II Ongoing
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: Versions 2.9.0 through 2.11.2 Product Description...
Reason: A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app...
Z-2164-2026 May 20, 2026 San Diego, CA
Technological Medical Advancements LLC
Class II Ongoing
Diowave Laser System, REF: Diowave 250W
Reason: Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthri...
Z-2162-2026 May 20, 2026 West Palm Beach, FL
DJO, LLC
Class II Ongoing
Brand Name: DONJOY Product Name: ICEMAN, W/COLDPAD, NS, RH Model/Catalog Number: 11-9099 Software Version: No Product Description: 11-9099 - ICEMAN, W/COLDPAD, NS, RH The IceMan CLASSIC3¿ cold t...
Reason: Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
Z-2137-2026 May 20, 2026 Carlsbad, CA
Dexcom, Inc.
Class II Ongoing
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: Versions 2.9.0 through 2.11.2 Product Description: De...
Reason: A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app...
Z-2163-2026 May 20, 2026 San Diego, CA
Mint Medical GmbH
Class II Ongoing
Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;
Reason: If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly li...
Z-2068-2026 May 13, 2026 Dossenheim
Intuitive Surgical, Inc.
Class II Ongoing
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical syst...
Reason: Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
Z-2046-2026 May 13, 2026 Sunnyvale, CA
Spacelabs Healthcare, Inc.
Class II Ongoing
Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A Product Description: The Command Module (91496) is the core of t...
Reason: Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Z-2052-2026 May 13, 2026 Snoqualmie, WA
Hologic, Inc
Class II Ongoing
Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes...
Reason: It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settin...
Z-1949-2026 May 6, 2026 Marlborough, MA
Boston Scientific Corporation
Class I Ongoing
Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U2...
Reason: Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 an...
Z-1771-2026 May 6, 2026 Saint Paul, MN
WHILL, INC.
Class I Ongoing
WHILL Model C2, with software versions before HMI: 2.22 / MC: 1.13; device is an indoor/outdoor battery-operated 2-wheel drive (rear-wheel drive) powered wheelchair.
Reason: Cyber vulnerability of BLE communication was found by CISA VDP program.
Z-1826-2026 May 6, 2026 Shinagawa
North American Rescue LLC.
Class I Ongoing
AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at th...
Reason: Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the ...
Z-1947-2026 May 6, 2026 Greer, SC
Advanced Bionics, LLC
Class II Ongoing
Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-550...
Reason: The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell batter...
Z-1950-2026 May 6, 2026 Valencia, CA
Beta Bionics, Inc.
Class II Ongoing
iLet Bionic Pancreas, REF: BB1001
Reason: insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, le...
Z-1996-2026 May 6, 2026 Irvine, CA
WHILL, INC.
Class I Ongoing
WHILL Model F, with software versions before HMI: 2.23 / MC: 1.05; the device is an indoor/outdoor, foldable battery-operated 2-wheel drive (rear-wheel drive) powered wheelchair.
Reason: Cyber vulnerability of BLE communication was found by CISA VDP program.
Z-1827-2026 May 6, 2026 Shinagawa
Boston Scientific Corporation
Class I Ongoing
Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100; 2. ESSENTIO DR SL Pacemaker, Model ...
Reason: Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 an...
Z-1770-2026 May 6, 2026 Saint Paul, MN
Zimmer, Inc.
Class II Ongoing
Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Comp...
Reason: Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromise...
Z-1946-2026 May 6, 2026 Warsaw, IN
Integra LifeSciences Corp. (NeuroSciences)
Class II Ongoing
CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.
Reason: Software issue that renders the touch screen unresponsive.
Z-1828-2026 April 29, 2026 Princeton, NJ
Foundation Medicine, Inc.
Class II Ongoing
Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx) installed on Research Triangle Park laboratory Equipment IDs 01342, 01591, 01989, 01990...
Reason: Potential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Hamilton STAR.
Z-1937-2026 April 29, 2026 Cambridge, MA
PHILIPS MEDICAL SYSTEMS
Class II Ongoing
Philips Spectral CT on Rails. Model Number: 728334.
Reason: Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing...
Z-1825-2026 April 22, 2026 Cambridge, MA
Gentuity, LLC
Class II Ongoing
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Reason: The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This...
Z-1823-2026 April 22, 2026 Sudbury, MA