Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

2,985 results found for software Page 8 of 120
ICU Medical, Inc.
Class II Ongoing
LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02
Reason: For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate meth...
Z-0965-2026 December 31, 2025 Lake Forest, IL
Zimmer, Inc.
Class II Ongoing
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509400 Software Version: N/A Product Description: Affixus 9 mm, Length 400 m...
Reason: Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occ...
Z-0971-2026 December 31, 2025 Warsaw, IN
Rocket Medical Plc
Class II Ongoing
Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54544-18-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with ...
Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Z-0952-2026 December 24, 2025 Tyne & Wear
Abbott Laboratories
Class II Ongoing
Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20
Reason: Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance i...
Z-0961-2026 December 24, 2025 Irving, TX
Mindray DS USA, Inc. dba Mindray North America
Class II Ongoing
Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model Number: 115-050935-00 and E115-050935-00. Software Version: ...
Reason: When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.
Z-0920-2026 December 24, 2025 Mahwah, NJ
GE Healthcare GmbH
Class II Ongoing
GE HealthCare ViewPoint 6, Catalog Number H47601AA; Radiological image processing system
Reason: GE HealthCare has become aware that the Patient History tab in ViewPoint 6 versions 6.15.3 and 6.15.4 ultrasound software may not include complete patient medical history informati...
Z-0935-2026 December 24, 2025 Munich
Rocket Medical Plc
Class II Ongoing
Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter with Safety Guard Model/Catalog Number: R54544-18-SG Software Version: N/A Product Description: Chest tube manufactured from PVC, wi...
Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Z-0953-2026 December 24, 2025 Tyne & Wear
Rocket Medical Plc
Class II Ongoing
Brand Name: Rocket Product Name: Rocket 16Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54549-16-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with ...
Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Z-0954-2026 December 24, 2025 Tyne & Wear
Rocket Medical Plc
Class II Ongoing
Brand Name: Rocket Product Name: Rocket 12Fg Seldinger Catheter with Safety Guard Model/Catalog Number: R54544-12-SG Software Version: N/A Product Description: Chest tube manufactured from PVC, wi...
Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Z-0951-2026 December 24, 2025 Tyne & Wear
DICOM Grid, Inc.
Class II Ongoing
Intelerad InteleShare software, with ProViewer component
Reason: Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare...
Z-0941-2026 December 24, 2025 Raleigh, NC
Fresenius Kabi USA, LLC
Class I Ongoing
Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.
Reason: Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.
Z-0885-2026 December 24, 2025 North Andover, MA
Fresenius Medical Care Holdings, Inc.
Class II Ongoing
5008X CAREsystem +CLiC +CDX;
Reason: Several software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled.
Z-0936-2026 December 24, 2025 Waltham, MA
Medtronic MiniMed, Inc.
Class II Ongoing
InPen App, Model/CFN Number: MMT-8061 (Android Users)
Reason: When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to...
Z-0958-2026 December 24, 2025 Northridge, CA
Rocket Medical Plc
Class II Ongoing
Brand Name: Rocket Product Name: Rocket 20Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54549-20-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with ...
Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Z-0955-2026 December 24, 2025 Tyne & Wear
Philips North America Llc
Class II Ongoing
CT 5300; Software Version Number: 4.5, 5.0, 5.1;
Reason: If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components locat...
Z-0882-2026 December 10, 2025 Cambridge, MA
Philips North America Llc
Class II Ongoing
Incisive CT; Software Version Number: 4.5, 5.0, 5.1;
Reason: If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components locat...
Z-0881-2026 December 10, 2025 Cambridge, MA
INTELERAD MEDICAL SYSTEMS INCORPORATED
Class II Ongoing
IntelePACS - InteleConnect / TechPortal
Reason: Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant ...
Z-0723-2026 December 3, 2025 Montreal, N/A
Medtronic MiniMed, Inc.
Class II Ongoing
CareLink Clinic, REF: MMT-7350
Reason: Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed ...
Z-0594-2026 December 3, 2025 Northridge, CA
DERMASENSOR INC
Class II Ongoing
DermaSensor. Model/Catalog Number: 10101. Software Version: 3.0.2. Software-aided adjunctive diagnostic device for use by physicians on lesions suspicious for skin cancer
Reason: Potential for device to not meet specifications resulting in incorrect results or delay of referral for patients.
Z-0583-2026 December 3, 2025 Miami, FL
Quest International, Inc.
Class II Ongoing
Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay Model/Catalog Number: 01-190 Software Version: N/A Product Description: Measles IgG tests, performed u...
Reason: Due to distributing the measles IgG IVD without a premarket approved/cleared.
Z-0534-2026 November 26, 2025 Doral, FL
Thoratec LLC
Class II Ongoing
Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS; 106524 Software V...
Reason: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Z-0505-2026 November 26, 2025 Pleasanton, CA
Beckman Coulter Inc.
Class II Ongoing
DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping up to ten fluorescent d...
Reason: Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
Z-0541-2026 November 26, 2025 Brea, CA
Thoratec LLC
Class II Ongoing
Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, H...
Reason: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Z-0508-2026 November 26, 2025 Pleasanton, CA
Exactech, Inc.
Class II Ongoing
Brand Name: Equinoxe Ergo Product Name: Ergo Augments Pilot Tip Converter Model/Catalog Number: 301-11-11 Software Version: N/A Product Description: Pilot Tip Converter Piece Component: Not a com...
Reason: Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Z-0537-2026 November 26, 2025 Gainesville, FL
Exactech, Inc.
Class II Ongoing
Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm Model/Catalog Number: 321-59-46 Software Version: N/A Product Description: rTSA Reamer, 46mm Component: Not a component
Reason: Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Z-0536-2026 November 26, 2025 Gainesville, FL