Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

2,985 results found for software Page 70 of 120
COVIDIEN LLC
Class I Terminated
Puritan Bennett 980 (PB980) ventilator series: Puritan Bennett" 980 Pediatric Adult Ventilator Puritan Bennett" 980 Neonatal Ventilator Puritan Bennett" 980 Universal Ventilator Product Usage:...
Reason: Software Update: External USB Drive performance and its impact on Graphic User Interface (GUI) functionality and labeling of the scalar waveform displayed on GUI during ventilation
Z-0497-2019 December 26, 2018 Mansfield, MA
GE Healthcare, LLC
Class II Terminated
Centricity Universal Viewer Zero Footprint Client, Model Number 2090258-001, Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2, SP9 and SP9.0.1 display o orientat...
Reason: Issue #1 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2. When an image is flipped horizontally and vertically, the orientation mar...
Z-0641-2019 December 26, 2018 Waukesha, WI
Ortho-Clinical Diagnostics
Class II Terminated
VITROS 5,1 FS Chemistry System, Version 3.0 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the an...
Reason: Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
Z-0651-2019 December 26, 2018 Rochester, NY
Siemens Healthcare Diagnostics, Inc.
Class II Terminated
Dimension Vista 1500, Model No. 10444801 an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to dete...
Reason: Siemens Healthcare Diagnostics has identified an issue with Dimension Vista software v.3.7 and v.3.8.1 that may result in aliquot well contamination. If a test is cancelled by the...
Z-0617-2019 December 26, 2018 Newark, DE
Ortho-Clinical Diagnostics
Class II Terminated
VITROS 5600 Integrated System, Version 3.3.1 & below For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITRO...
Reason: Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
Z-0650-2019 December 26, 2018 Rochester, NY
bioMerieux, Inc.
Class II Terminated
VITEK¿ 2 Systems Software Version 9.01 Update Kit.
Reason: The VITEK¿ 2 Systems Software Version 9.01 software may display incorrect organism identification information as well as the associated Antimicrobial Susceptibility Testing (AST) ...
Z-0639-2019 December 26, 2018 Durham, NC
PerkinElmer Health Sciences Canada, Inc.
Class II Terminated
PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Pa...
Reason: Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acqu...
Z-0513-2019 December 5, 2018 Bolton, N/A
Synaptive Medical Inc
Class I Terminated
BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s cus...
Reason: This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is s...
Z-0470-2019 December 5, 2018 Toronto, N/A
Beckman Coulter Inc.
Class I Terminated
1) Cytomics FC 500 With CXP Software BECKMAN COULTER 2) Cytomics FC 500 MPL With MXP Software BECKMAN COULTER flow cytometer for in vitro diagnostics
Reason: A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plot...
Z-0471-2019 November 28, 2018 Brea, CA
Tosoh Bioscience Inc
Class II Terminated
Tosoh AIA-900 immunoassay Analyzer
Reason: A software issue may cause the liquid volume detection entries for 2nd and 3rd reagent bottles to be overwritten with the data from the 1st reagent bottle. Tosoh has released softw...
Z-0485-2019 November 28, 2018 Grove City, OH
Beckman Coulter Inc.
Class I Terminated
COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software
Reason: A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plot...
Z-0472-2019 November 28, 2018 Brea, CA
Invivo Corporation
Class II Completed
Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388
Reason: The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or...
Z-0462-2019 November 21, 2018 Orlando, FL
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class II Ongoing
Medtronic CareLink Encore 29901 Programmer
Reason: Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an ...
Z-0407-2019 November 14, 2018 Mounds View, MN
Medtronic Navigation, Inc.-Littleton
Class II Terminated
Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000...
Reason: O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new...
Z-0404-2019 November 14, 2018 Littleton, MA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class II Ongoing
Medtronic CareLink 2090 Programmer
Reason: Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an ...
Z-0406-2019 November 14, 2018 Mounds View, MN
Philips Medical Systems (Cleveland) Inc
Class II Terminated
Brilliance Big Bore Radiology CT (Model No. 728244) software version 4.2.0 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer recon...
Reason: It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provid...
Z-0307-2019 November 7, 2018 Cleveland, OH
RAYSEARCH LABORATORIES AB
Class II Terminated
RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7, a stand-alone software treatment planning system.
Reason: Robust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7.
Z-0294-2019 November 7, 2018 Stockholm, N/A
Beckman Coulter Inc.
Class III Terminated
Sorting-Drive Product Part, Sorting-Drive Plus Kit (Sorting Drive Software v4.2.1), Part No. A89483, Beckman Coulter, Inc. The Sorting-Drive Plus Kit includes the Sorting-Drive 4.2.1 User Kit whic...
Reason: Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. Sorting Dr...
Z-0358-2019 November 7, 2018 Brea, CA
Fresenius Vial Sa
Class II Terminated
Volumat MC Agilia Volumetric Infusion Pump, Model No. Z021135, Software version 1.7. Product Usage - The Volumat MC Agilia is a piece of transportable medical equipment intended to be used in healthca...
Reason: Fresenius Kabi initiated a mandatory software upgrade of the Volumat MC Agilia Volumetric Infusion Pump because it was designed to improve customer experience while preventing comm...
Z-0362-2019 November 7, 2018 Brezins, N/A
Beckman Coulter Inc.
Class III Terminated
Software Package, Sorting-Drive 4.2 User Kit, Part No. B73488, Manufactured by: Beckman Coulter Biomedical GmbH The Sorting-Drive itself is an optional software to be used with the Automate 2500 F...
Reason: Sorting-Drive software version 4.2 (released in October 2015) introduced a new option which included the Sorting-Drive Client that is used for manual printing of barcodes. When ...
Z-0354-2019 November 7, 2018 Brea, CA
Philips Medical Systems (Cleveland) Inc
Class II Terminated
Ingenuity Core (Model No. 728321), Software version 4.1.6 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ra...
Reason: It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provid...
Z-0309-2019 November 7, 2018 Cleveland, OH
Beckman Coulter Inc.
Class III Terminated
Sorting-Drive Professional Part, Sorting-Drive Professional Kit, (Sorting Drive Software v4.2.1), Part No. A94144, Beckman Coulter, Inc. The Sorting-Drive Professional Kit (PN A94144) includes t...
Reason: Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes.Sorting Dri...
Z-0359-2019 November 7, 2018 Brea, CA
Philips Medical Systems (Cleveland) Inc
Class II Terminated
Brilliance iCT SP (Model No. 728311, Software Version 4.7.0 or 4.7.2 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstructio...
Reason: It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provid...
Z-0311-2019 November 7, 2018 Cleveland, OH
Beckman Coulter Inc.
Class III Terminated
Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc. The iPAW ((intelligent Peri-Analytical Workstation) can act as: a software client f...
Reason: Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. Sorting Dr...
Z-0353-2019 November 7, 2018 Brea, CA
Philips Medical Systems (Cleveland) Inc
Class II Terminated
Vereos PET/CT, Software version 2.0.1. Model Number: 882446 Product Usage: The Philips Vereos PET/CT system is a diagnostic imaging device that combines Positron Emission Tomography (PET) and X...
Reason: Software Correction: Software errors that may result in incorrect 4D CT images, PET images with SUV quantification error, or disruption of interventional CT procedures; and Softwar...
Z-0302-2019 November 7, 2018 Cleveland, OH