RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6) . Intended Use: software system designed for treatm...
Reason: When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong
Z-1930-2019July 10, 2019Stockholm, N/A
GE Healthcare, LLC
Class IITerminated
Centricity PACS Foundation Centricity PACS software product is intended for the storage, reading, diagnostic review, analysis, annotation, distribution, printing, editing, and processing of digital...
Reason: There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the arch...
Z-1911-2019July 10, 2019Waukesha, WI
Siemens Medical Solutions USA, Inc
Class IITerminated
SOMATOM go.Now, Model No. 11061618 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional im...
Reason: The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.
Z-1920-2019July 10, 2019Malvern, PA
Siemens Medical Solutions USA, Inc
Class IITerminated
SOMATOM go.All (Model #11061630), with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images...
Reason: The potential sporadic performance problems may result in scanning workflow interruptions and unexpected user notifications. This may cause a delay in diagnosis or patient scans.
Z-1917-2019July 10, 2019Malvern, PA
Siemens Medical Solutions USA, Inc
Class IITerminated
SOMATOM go.Top, Model No. 11061648 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional im...
Reason: The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.
Z-1922-2019July 10, 2019Malvern, PA
Siemens Medical Solutions USA, Inc
Class IITerminated
SOMATOM go.All, Model No. 11061638 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional im...
Reason: The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.
Z-1921-2019July 10, 2019Malvern, PA
Siemens Medical Solutions USA, Inc
Class IITerminated
SOMATOM go.Now (Model #11061620), with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images...
Reason: The potential sporadic performance problems may result in scanning workflow interruptions and unexpected user notifications. This may cause a delay in diagnosis or patient scans.
Z-1916-2019July 10, 2019Malvern, PA
Siemens Medical Solutions USA, Inc
Class IITerminated
SOMATOM go.Up (Model #11061610) with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of...
Reason: The potential sporadic performance problems may result in scanning workflow interruptions and unexpected user notifications. This may cause a delay in diagnosis or patient scans.
Z-1915-2019July 10, 2019Malvern, PA
Hitachi America, Ltd., Power Systems Division
Class IITerminated
PROBEAT-V, software versions 4.8.1S, and 5.1M Radiation therapy system.
Reason: There is a potential for a discrepant target position when using 3D3D matching mode in PIAS (Positioning Image Analysis System) software installed in the PROBEAT-V, proton therapy ...
Z-1892-2019July 3, 2019Houston, TX
Arkray Factory USA, Inc.
Class IITerminated
AUTION HYBRID AU-4050
Reason: This correction is being initiated due to a software issue which results in the possibility of incorrect patient information being assigned to sample results when the following rar...
Z-1850-2019June 26, 2019Minneapolis, MN
Philips North America, LLC
Class IIITerminated
Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator: 866389 Patient Information Center iX B.01 (or later); 866390 Patient Information ...
Reason: Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Cen...
Z-1771-2019June 19, 2019Andover, MA
Flowonix Medical Inc
Class IITerminated
Prometra Programmable Pump, Cat. No. 11827 For intrathecal infusion.
Reason: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a softwa...
Z-1714-2019June 12, 2019Mansfield, MA
Fresenius Medical Care Renal Therapies Group, LLC
Class IIITerminated
2008T BlueStar Machine Software v.2.72-Hemodialysis machine Model Number: 191126 - Product Usage: The 2008T hemodialysis machine is indicated for acute and chronic dialysis therapy in a healthcare f...
Reason: Power Logic Board Fail Message on the 2008T BlueStar Machine with software version 2.72.
Reason: Updated Warning Statements in the Operator's Manual/Directions for Use and Software Updates to reduce the risk of potential patient harm due to a delay in performing a procedure
Reason: Updated Warning Statements in the Operator's Manual/Directions for Use and Software Updates to reduce the risk of potential patient harm due to a delay in performing a procedure
Z-1702-2019June 12, 2019Glens Falls, NY
Flowonix Medical Inc
Class IITerminated
Prometra II Programmable Pump (Clinical), Cat. No. 01827 For intrathecal infusion.
Reason: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a softwa...
Reason: Updated Warning Statements in the Operator's Manual/Directions for Use and Software Updates to reduce the risk of potential patient harm due to a delay in performing a procedure
Z-1703-2019June 12, 2019Glens Falls, NY
Flowonix Medical Inc
Class IITerminated
Prometra Clinician Programmer (Touchscreen) Software version 2.00.29, Cat. No. 12828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.
Reason: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a softwa...
Reason: Updated Warning Statements in the Operator's Manual/Directions for Use and Software Updates to reduce the risk of potential patient harm due to a delay in performing a procedure
Reason: Updated Warning Statements in the Operator's Manual/Directions for Use and Software Updates to reduce the risk of potential patient harm due to a delay in performing a procedure
Z-1704-2019June 12, 2019Glens Falls, NY
Flowonix Medical Inc
Class IITerminated
Prometra II Programmable Pump, Cat. No. 13827 For intrathecal infusion.
Reason: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a softwa...
Z-1715-2019June 12, 2019Mansfield, MA
Flowonix Medical Inc
Class IITerminated
Prometra Programmer (Grand Prime) Software version 2.00.29, Cat. No. 13828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.
Reason: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a softwa...
Z-1713-2019June 12, 2019Mansfield, MA
Abbott Laboratories Inc. (St Jude Medical)
Class IITerminated
TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)
Reason: In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. ...
Z-1493-2019June 5, 2019Saint Paul, MN
Neusoft Medical Systems Co., Ltd.
Class IITerminated
NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube an...
Reason: Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system wh...
Z-1279-2019May 29, 2019Shenyang, N/A
Roche Diagnostics Corporation
Class IITerminated
Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Data Processing Module, For Clinical Use
Reason: Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when usin...