VITROS 3600 Immunodiagnostic System Software Versions 3.3.3 and below Product Code: 6802783, Unique Identifier: 10758750002979
Reason: Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances
Z-2480-2020July 8, 2020Rochester, NY
Ortho Clinical Diagnostics
Class IITerminated
VITROS XT 7600 Integrated System Software Versions 3.5.1 and below Product Code:6844461 Unique Identifier: 10758750031610
Reason: Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances
Z-2484-2020July 8, 2020Rochester, NY
Fujifilm Medical Systems U.S.A., Inc.
Class IIOngoing
Fujifilm Synapse PACS Software Version 5.5.x and 5.7.x- intended as interface to the FUJIFILM Synapse PACS Software (Server)
Reason: Lateromedial (LM) and Lateromedial Oblique (LMO) Orientation Markers may be displayed incorrectly on the 3D TOMO slider bar. In addition, the slice location of the image and/or the...
Z-2471-2020July 8, 2020Lexington, MA
Preventice Services, LLC
Class IITerminated
BodyGuardian Heart Remote Monitoring Kit, consisting of two BG Heart monitors, either Models BC-1001-A or BC1001-D, and a cell phone, either Models BP1008 or BP10014, that is paired to both of the mon...
Reason: The device data being collected and transferred to the monitoring center may not be accurate due to nonvalidated association between the phone software and the heart monitors, ther...
Z-2479-2020July 8, 2020Houston, TX
RAYSEARCH LABORATORIES AB
Class IIOngoing
RayStation standalone software treatment planning system (RayStation 8A, 8B, 9A, and 9B, including all service packs) model no. 8.0, 8.1, 9.0, 9.1, 9.2
Reason: It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found wit...
Z-2426-2020July 1, 2020Stockholm, N/A
Merge Healthcare, Inc.
Class IITerminated
Merge Healthcare Merge Application Server Software Release and Merge Healthcare Merge Cardio Workstation Software Release - Product Usage: VERICIS is a system intended to be used to acquire, store, pr...
Reason: An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abn...
Z-2403-2020July 1, 2020Hartland, WI
Abaxis Inc
Class IITerminated
Piccolo Xpress chemistry analyzer, Model No. 1100-1001, Software version 2.1.55 (NGA)/3.1.35(Jen II)
Reason: In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitor...
Z-2297-2020June 17, 2020Crawley, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 8B, UDI # 07350002010129 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2346-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 6 Service Pack 1, UDI #s 07350002010037, 07350002010082 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2352-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 9B Service Pack 1, UDI # 07350002010297 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2351-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 8A, UDI # 07350002010112 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2344-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 9B, UDI # 07350002010266 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2350-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 7, UDI # 07350002010068 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2353-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 4.0 to RayStation 5 Service Pack 2 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2339-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 9A, UDI # 07350002010174 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2349-2020June 17, 2020Stockholm, N/A
Becton, Dickinson and Company, BD Biosciences
Class IITerminated
BD FACS Sample Prep Assistant (SPA) II - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto clinical software when used with the BD FACS Sample Prep Assistant II /III is intended for us...
Reason: The recalling firm confirmed that BD FACS Sample Prep Assistant (SPA) II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline i...
Z-2356-2020June 17, 2020San Jose, CA
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 6 Service Pack 3, UDI # 07350002010242 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2343-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 5 Service Pack 3, UDI # 07350002010020 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2340-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 8B Service Pack 1, UDI # 07350002010204 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2347-2020June 17, 2020Stockholm, N/A
Becton, Dickinson and Company, BD Biosciences
Class IITerminated
BD FACS Sample Prep Assistant (SPA) II Upgrade - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto clinical software when used with the BD FACS Sample Prep Assistant II /III is intende...
Reason: The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for...
Z-2357-2020June 17, 2020San Jose, CA
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 6, UDI # 07350002010013 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2341-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 6 Service Pack 2, UDI # 07350002010075 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2342-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 8B Service Pack 2, UDI # 07350002010235 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Z-2348-2020June 17, 2020Stockholm, N/A
RAYSEARCH LABORATORIES AB
Class IITerminated
RayStation 8A Service Pack 1, UDI # 07350002010136 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.