BIOFLO 8F SINGLE TITANIUM NON- FILLED; VALVED, UPN H965440250, Cat. No. 44-025 - Product Usage: Long-term access to the central venous system for administration of fluids including but not limited to ...
Reason: Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connec...
Z-0502-2020November 27, 2019Glens Falls, NY
Neocis Inc.
Class IITerminated
Neocis Guidance System Fiducial Array - Product Usage: The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and th...
Reason: The Fiducial Array may mismatch their measurement files resulting in failed landmark check.
Z-0511-2020November 27, 2019Miami, FL
Angiodynamics Inc. (Navilyst Medical Inc.)
Class IITerminated
BIOFLO 8F SINGLE PLASTIC FILLED; NON-VALVED, UPN H965440140, Cat. No. 44-014 - Product Usage: Long-term access to the central venous system for administration of fluids including but not limited to hy...
Reason: Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connec...
Z-0501-2020November 27, 2019Glens Falls, NY
Angiodynamics Inc. (Navilyst Medical Inc.)
Class IITerminated
BIOFLO 8F SINGLE PLASTIC NON-FILLED; VALVED, UPN H965440210, Cat. No. 44-021 - Product Usage: Long-term access to the central venous system for administration of fluids including but not limited to hy...
Reason: Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connec...
Z-0503-2020November 27, 2019Glens Falls, NY
Implant Direct Sybron Manufacturing, LLC
Class IITerminated
Implant Direct, REF: 833708, Legacy 2 Implant, Surface: HA, 3.7mmD x 8mmL, 3.5mmD Platform, RxOnly, Sterile R, Made in USA Dental Implants
Reason: A QA inspection confirmed that a Healing Collar (Part number 8735-15) was packaged instead of the Implant ordered.
Reason: packaging error; An internal QA inspection revealed that a incorrect healing collar could have been packaged in an implant assembly kit.
Z-0363-2020November 20, 2019Westlake Village, CA
Exactech, Inc.
Class IITerminated
2.4mm x 2.5 Fluted Pin Pouch; Catalog Number: 351-90-22; Orthopedic surgical instrument used in the implantation of the Vantage Total Ankle System - Product Usage: To assist the surgeon in the implan...
Reason: Potential for the 2.4mm x 2.5 Fluted Pin Pouch to contain a single pin rather than four (4) as intended.
Z-0361-2020November 20, 2019Gainesville, FL
Zimmer Biomet, Inc.
Class IITerminated
SM DIS VOL/DORS RAD LAT COL PLATE, Item Nos. 47235801404 47235802102 47235802103 47235802104 47235802106 47235802108 47235802203 47235802204 47235802206 47235802208 47235802210
Reason: Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Reason: Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Z-0325-2020November 13, 2019Warsaw, IN
Zimmer Biomet, Inc.
Class IITerminated
Various Surgical Instruments: Screws and Reamers, Item Nos. 00579104100 00579104200 00579104300 00579104400 00579105400 00597909526 00597909529 00597909532 00597909535 00597909538 0059790954...
Reason: Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Z-0322-2020November 13, 2019Warsaw, IN
LivaNova USA Inc
Class IIOngoing
The LivaNova VNS Therapy System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. Th...
Reason: Lead impedance values reported by the affected VNS generator will be higher compared to those reported by previous models. This is due to a change in the timing of when affected VN...