Abbott
Class II Terminated
TactiCath Quartz Contact Force Ablation Catheter, Model Numbers: PN-004 065 (GTIN 07640157990033) and PN-004 075 (GTIN 07640157990040) - Product Usage: is a multi-electrode irrigated catheter with a ...
Reason: High rate of reported complaints for the loss of contact force information while using the TactiCath Quartz Contact Force Ablation Catheter
Z-2541-2020 July 15, 2020 Plymouth, MN
Smiths Medical ASD Inc.
Class II Terminated
PORT-A-CATH¿ II POWER P.A.C. Low Profile" system.
Reason: Smiths Medical became aware that one (1) lot of PORT-A-CATH¿ II POWER P.A.C. Low Profile" System (SKU 21-4477-24) was packaged with an 8Fr catheter instead of a 6Fr catheter.
Z-2499-2020 July 15, 2020 Minneapolis, MN
Bound Tree Medical
Class II Terminated
Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the...
Reason: Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supr...
Z-2532-2020 July 15, 2020 Dublin, OH
Prodimed Plastimed Division
Class II Terminated
Prodimed Combicath catheters for blind or fibrescope-guided protected distal bronchial sampling (PDBS)- COMBICATH PE 90 cm Product Code: 58229 19
Reason: Lack of assurance of sterility
Z-2490-2020 July 15, 2020 Le Plessis Bouchard
Prodimed Plastimed Division
Class II Terminated
Prodimed Combicath catheters for mini bronchoalveolar lavage (BAL)-COMBICATH PE 60 cm x 2.7 mm Product Code: 58216 27US
Reason: Lack of assurance of sterility
Z-2487-2020 July 15, 2020 Le Plessis Bouchard
Prodimed Plastimed Division
Class II Terminated
Prodimed Combicath Paediatric catheters for blind protected distal bronchial sampling- COMBICATH 35 cm x 1.9 mm Product Code: 58 223 19US
Reason: Lack of assurance of sterility
Z-2491-2020 July 15, 2020 Le Plessis Bouchard
Prodimed Plastimed Division
Class II Terminated
Prodimed Combicath catheters for blind or fibrescope-guided protected distal bronchial sampling (PDBS)- COMBICATH 90 cm x 1.9 mm Product Code: 58229 19US
Reason: Lack of assurance of sterility
Z-2489-2020 July 15, 2020 Le Plessis Bouchard
Angiodynamics, Inc.
Class II Terminated
BioFlo Midline Catheter and BioFlo Maximal Barrier Kits - Product Usage: These introducers are used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters or pacing ...
Reason: Certain AngioDynamics PICC Kits contain BD PosiFlush SF Saline Flush Syringes which were recalled by BD Medical on April 16, 2020.
Z-2395-2020 June 24, 2020 Queensbury, NY
BIOSENSE WEBSTER
Class II Terminated
PENTARAY NAV HIGH-DENSITY MAPPING CATHETER ELECTRODE EA 22 Rx only STERILE EO - Product Usage: The Biosense Webster PENTARAY NAV High-Density Mapping Catheter is indicated for multiple electrode ...
Reason: Mapping Catheter can become entrapped in the cardiac valvular apparatus while the user is attempting to create an electroanatomic map.
Z-2371-2020 June 24, 2020 IRVINE, CA
BIOSENSE WEBSTER
Class II Terminated
PENTARAY NAV eco HIGH-DENSITY MAPPING CATHETER ELECTRODE EA 22 Rx only STERILE EO - Product Usage: The Biosense Webster PENTARAY NAV High-Density Mapping Catheter is indicated for multiple electr...
Reason: Mapping Catheter can become entrapped in the cardiac valvular apparatus while the user is attempting to create an electroanatomic map.
Z-2372-2020 June 24, 2020 IRVINE, CA
Angiodynamics, Inc.
Class II Terminated
BioFlo Midline Convenience Kits - Product Usage: These introducers are used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters or pacing leads in the vasculature...
Reason: Certain AngioDynamics PICC Kits contain BD PosiFlush SF Saline Flush Syringes which were recalled by BD Medical on April 16, 2020.
Z-2396-2020 June 24, 2020 Queensbury, NY
Canon Medical System, USA, INC.
Class II Ongoing
Infinix-8000F with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration...
Reason: Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
Z-2327-2020 June 17, 2020 Tustin, CA
Arrow International Inc
Class II Ongoing
Arrow FlexTip Plus Epidural Kit with NRFit Connector, catheter system, Product Codes (US) OT19TAGSK OT19TAGSKA OT19TAPSK OT19TCGLK Product Codes (OUS) AT-05501-NRF OT19HBPSS UM-05400-NRF AS...
Reason: Nine complaints reported that the rotatable collar on the filter had detached, causing a leak.
Z-2288-2020 June 17, 2020 Reading, PA
INARI MEDICAL
Class II Terminated
Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI: 00850291007079
Reason: The hemostasis valve of aspiration guide catheter devices may fail to seal resulting in a loss of vacuum during thrombus aspiration, which may result the procedure being prolonged ...
Z-2299-2020 June 17, 2020 Irvine, CA
Canon Medical System, USA, INC.
Class II Ongoing
Infinix-8000C with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration...
Reason: Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
Z-2326-2020 June 17, 2020 Tustin, CA
Canon Medical System, USA, INC.
Class II Ongoing
Infinix-8000V with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration...
Reason: Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
Z-2325-2020 June 17, 2020 Tustin, CA
Bard Peripheral Vascular Inc
Class II Terminated
EQUISTREAM¿ XK Long-Term Hemodialysis Catheter. REF/UDI:5913190/(01)00801741013836; 16F
Reason: Incorrect Peel-Apart Introducer Sheath in Kit.
Z-2209-2020 June 10, 2020 Tempe, AZ
Bard Peripheral Vascular Inc
Class II Terminated
EQUISTREAM¿ XK Long-Term Hemodialysis Catheter. REF/UDI:5913230/:(01)00801741013843; 16F
Reason: Incorrect Peel-Apart Introducer Sheath in Kit.
Z-2208-2020 June 10, 2020 Tempe, AZ
Argon Medical Devices, Inc
Class II Terminated
Skater Introducer Set 6F x 18cm. REF/UDI: 651506300/(01)00886333005000; 655506300/(01)00886333005079 - Product Usage: The Skater Introducer System with radiopaque marker is intended for fine needle pu...
Reason: Tungsten coils of the guidewire included in the introducer kit were detaching.
Z-2109-2020 June 3, 2020 Athens, TX
Medtronic Vascular Galway DBA Medtronic Ireland
Class I Terminated
Medtronic Confida Expandable Sheath, REF CES122230. To provide a guide for catheters or devices introduced into the femoral iliac arteries.
Reason: Increased risk of events associated with vascular complications (bleeding, hematoma, and dissection)
Z-2039-2020 June 3, 2020 Galway, N/A
Medline Industries Inc
Class II Terminated
SELECTSILICONE" 100% SILICONE FOLEY CATHETER 2-WAY 8 Fr 3 ml - Product Usage: Medline Silicone Foley catheter is a flexible tubular device that is inserted through the urethra and used to drain urine ...
Reason: Loose silicone particulate was found to be present on the shaft of the silicone catheters.
Z-2043-2020 May 27, 2020 Northfield, IL
Medline Industries Inc
Class II Terminated
SELECTSILICONE" 100% SILICONE FOLEY CATHETER 2-WAY 6 Fr 1.5 ml - Product Usage: Medline Silicone Foley catheter is a flexible tubular device that is inserted through the urethra and used to drain urin...
Reason: Loose silicone particulate was found to be present on the shaft of the silicone catheters.
Z-2042-2020 May 27, 2020 Northfield, IL
Medline Industries Inc
Class II Terminated
SELECTSILICONE" 100% SILICONE FOLEY CATHETER 2-WAY 10 Fr 3 ml - Product Usage: Medline Silicone Foley catheter is a flexible tubular device that is inserted through the urethra and used to drain urine...
Reason: Loose silicone particulate was found to be present on the shaft of the silicone catheters.
Z-2044-2020 May 27, 2020 Northfield, IL
Covidien Llc
Class II Terminated
Dover 100% Silicone Foley Catheter with Temperature 16FR 5CC2W, Product Code: 50516T - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use ...
Reason: Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could ...
Z-1986-2020 May 20, 2020 Mansfield, MA
Edwards Lifesciences, LLC
Class II Terminated
Sterile Oxymetry Oligon Catheter Kit. Pressure injectable. REF/UDI:XA3816SAFKT/ (01)00690103197181
Reason: Sterile catheter kit packaging compromised.
Z-2007-2020 May 20, 2020 Irvine, CA