Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

2,985 results found for software Page 45 of 120
Siemens Medical Solutions USA, Inc
Class II Terminated
syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgr...
Reason: Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed i...
Z-0995-2022 May 4, 2022 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Terminated
syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Sim Model Number 11061660. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgr...
Reason: Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed i...
Z-0996-2022 May 4, 2022 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Terminated
syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Open Pro Model Number 11061670. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A afte...
Reason: Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed i...
Z-0997-2022 May 4, 2022 Malvern, PA
Roche Diabetes Care, Inc.
Class II Terminated
RocheDiabetes Care Platform Software version 2.5.2 material number 08888353001. This is a web application software hosted on a server. There is no physical product.
Reason: Potential for patient data mismatch when using browser "back" button to navigate between patients when using the diabetes management software.
Z-1008-2022 May 4, 2022 Indianapolis, IN
Medtronic Neuromodulation
Class II Ongoing
MEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL, which includes the following: a) Handset with Communicator, Model Numbers: 1) TH90Q01, 2) TH90QFA, 3) TH90PFA, 4) TH90GFA, 5) TG90G01; b) Inte...
Reason: There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost"...
Z-0998-2022 May 4, 2022 Minneapolis, MN
Baxter Healthcare Corporation
Class II Ongoing
PrisMax System, Product Codes: a) 955558, b) 955627, c) 955701
Reason: This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limi...
Z-0993-2022 May 4, 2022 Deerfield, IL
Baxter Healthcare Corporation
Class II Ongoing
PrisMax System, Product Code 955626
Reason: This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limi...
Z-0991-2022 May 4, 2022 Deerfield, IL
Siemens Medical Solutions USA, Inc
Class II Terminated
syngo.via RT Image Suite upgraded to VB60 running on syngo.via Model Number 10496180. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
Reason: Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed i...
Z-0994-2022 May 4, 2022 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Ongoing
Artis icono floor in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327700
Reason: Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential softwar...
Z-0986-2022 April 27, 2022 Malvern, PA
Medtronic Xomed, Inc.
Class II Terminated
Software 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND EHS, Product Number E...
Reason: As a result of Medtronic's integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect ...
Z-0961-2022 April 27, 2022 Jacksonville, FL
Siemens Medical Solutions USA, Inc
Class II Ongoing
Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Reason: Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential softwar...
Z-0987-2022 April 27, 2022 Malvern, PA
St. Jude Medical, Cardiac Rhythm Management Division
Class II Ongoing
Merlin 2 PCS MER3700 programmer Model MER3400 software
Reason: Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchroniza...
Z-0965-2022 April 27, 2022 Sylmar, CA
St. Jude Medical, Cardiac Rhythm Management Division
Class II Ongoing
Merlin PCS 3650 programmer Model 3330 software
Reason: Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchroniza...
Z-0964-2022 April 27, 2022 Sylmar, CA
Siemens Medical Solutions USA, Inc
Class II Ongoing
SIEMENS SOMATOM with software syngo CT VA30A_SP4 Models: SOMATOM go.Now 11061610 & 11061618 SOMATOM go.Up 11061620 & 11061628 SOMATOM go.All 11061630 & 11061638 SOMATOM go.Top 11061640 &...
Reason: Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur d...
Z-0962-2022 April 27, 2022 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Ongoing
Artis icono biplane- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327600
Reason: Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential softwar...
Z-0985-2022 April 27, 2022 Malvern, PA
SIGHT DIAGNOSTICS LTD
Class II Ongoing
The Sight OLO device is a computer vision based platform for blood analysis. The platform combines computer-vision algorithms for image processing to identify and quantify blood components (e.g., re...
Reason: In instances where custom reference ranges were configured on the device post installation, a possibility to inadvertently apply changes to reference range values was found on sof...
Z-0963-2022 April 27, 2022 Tel Aviv-Yafo
Varian Medical Systems Imaging Laboratory GmbH
Class II Completed
ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1
Reason: Software issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location
Z-0942-2022 April 20, 2022 Baden, N/A
Volcano Corporation
Class II Ongoing
SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x
Reason: If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be dis...
Z-0885-2022 April 13, 2022 Rancho Cordova, CA
Philips North America Llc
Class II Ongoing
Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224,...
Reason: Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown ...
Z-0874-2022 April 13, 2022 Cambridge, MA
Philips Healthcare
Class II Terminated
Philips StentBoost Live R2.0 application, when used with the Philips Azurion with software release R2.x
Reason: Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live pr...
Z-0737-2022 April 13, 2022 Andover, MA
Remote Diagnostic Technologies Ltd.
Class II Terminated
Tempus LS, Type DGE1, CAT 00-3010, REF 3.940590, Software version 1.3.5, UDI: (01)07613365001693
Reason: Distribution of Defibrillators that are not approved or cleared for distribution US Market.
Z-0861-2022 April 13, 2022 Basingstoke
Draeger Medical, Inc.
Class II Terminated
Evita V500 Ventilator with Software 2.51.01 and Lower - intended for the ventilation of adult, pediatric and neonatal patients Catalog Number: 8416400
Reason: Cybersecurity Improvement Action-not equipped against potential cyber security threats
Z-0870-2022 April 13, 2022 Telford, PA
Draeger Medical, Inc.
Class II Terminated
Babylog VN500 Ventilator with Software 2.51.01 and Lower- intended for the ventilation neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 ...
Reason: Cybersecurity Improvement Action-not equipped against potential cyber security threats
Z-0871-2022 April 13, 2022 Telford, PA
Diagnos Inc
Class II Terminated
CARA (Computer Assisted Retinal Analysis), Product ID: +B543CARA0/$$+7V4.7.2/16D202111179
Reason: Retinal analysis system software has some features, based on artificial intelligence, that potentially lack appropriate market approval, so it is recommended to use a specific prio...
Z-0826-2022 April 6, 2022 Brossard, N/A
RaySearch America Inc
Class II Ongoing
RayStation 4-11B, and RayPlan 1, 2, 7-11B including some service packs: RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35), RayStation 5 Service Pack 3 (5.0.3.17), RayStation 6...
Reason: Incorrect source to skin or surface distance (SSD) calculation may result in dose error when using RayStation system.
Z-0850-2022 April 6, 2022 New York, NY