Reason: Sterility may be compromised due to an unsealed pouch
Z-1245-2021March 17, 2021Danvers, MA
Medtronic Vascular
Class IITerminated
Medtronic 6F Launcher Guide Catheter 100 cm JR 4.0 Model Number: LA6JR40
Reason: Sterility may be compromised due to an unsealed pouch
Z-1246-2021March 17, 2021Danvers, MA
Medtronic Vascular
Class IITerminated
Medtronic 6F Launcher Guide Catheter 100 cm EBU 3.0 Model Number: LA6EBU30
Reason: Sterility may be compromised due to an unsealed pouch
Z-1248-2021March 17, 2021Danvers, MA
Medical Components, Inc dba MedComp
Class IIOngoing
Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.
Reason: The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to ...
Z-1184-2021March 17, 2021Harleysville, PA
Medtronic Vascular
Class IITerminated
Medtronic 6F Launcher Guide Catheter 100cm EBU 3.75 Model Number: LA6EBU375
Reason: Sterility may be compromised due to an unsealed pouch
Z-1247-2021March 17, 2021Danvers, MA
Acist Medical Systems
Class ITerminated
ACIST Kodama Intravascular Ultrasound Catheter The Kodama Intravascular Ultrasound Catheter is a component of the ACIST HDi System. The ACIST HDi System is intended to be used for ultrasound exami...
Reason: Test results from the manufacturing line found a piece of damaged o-ring in an unexpected section of the catheter. Further testing indicated that pieces (>200 micron) of damaged o-...
Z-1161-2021March 10, 2021Eden Prairie, MN
Siemens Medical Solutions USA, Inc
Class IITerminated
Sensis / Sensis Vibe systems with software version VD12 and equipped with HP Flex Pro-C PC. Diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies...
Reason: System may sporadically freeze (lock-up) during operation or while being in an idle state, no longer possible to interact with the system
Z-1173-2021March 10, 2021Malvern, PA
Aomori Olympus Co., Ltd.
Class IITerminated
B5-2Q BALLOON CATHETER, Model No. B5-2Q - Product Usage: Used with endoscopes to inject contrast medium or other medical fluid into the bile duct, pancreatic duct, and respiratory organs.
Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Z-0961-2021February 17, 2021Kuroishi, N/A
ICU Medical, Inc.
Class IITerminated
60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave¿ Clear, 1.2 Micron Filter, 2 Clamps, Luer Lock REF MC330174 - Product Usage: The intravascular administration set is a single use, sterile, no...
Reason: Due to the incorrect filter assembly.
Z-0956-2021February 17, 2021San Clemente, CA
Aomori Olympus Co., Ltd.
Class IITerminated
B7-2C BALLOON CATHETER 13mm dia 2.8; model no. B7-2C - Product Usage: Used with endoscopes to inject contrast medium or other medical fluid into the bile duct, pancreatic duct, and respiratory organs.
Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Z-0963-2021February 17, 2021Kuroishi, N/A
Aomori Olympus Co., Ltd.
Class IITerminated
B7-2Q BALLOON CATHETER 13MM DIA 3.2, model no. B7-2Q -Product Usage: Used with endoscopes for injection of contrast medium or other medical fluid into the biliary or pancreatic tract.
Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Z-0965-2021February 17, 2021Kuroishi, N/A
Argon Medical Devices, Inc
Class IITerminated
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18S, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/6f Straight Cat...
Reason: Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could res...
Z-1072-2021February 17, 2021Athens, TX
ICU Medical, Inc.
Class IITerminated
60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides access for th...
Reason: Due to the incorrect filter assembly.
Z-0955-2021February 17, 2021San Clemente, CA
Aomori Olympus Co., Ltd.
Class IITerminated
B5-2C BALLOON CATHETER 11MM DIA 2.0; Model No. B5-2C - Product Usage: Used with endoscopes to inject contrast medium or other medical fluid into the bile duct, pancreatic duct, and respiratory organs.
Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Z-0962-2021February 17, 2021Kuroishi, N/A
Argon Medical Devices, Inc
Class IITerminated
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19S, 1/19 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/5f straight ca...
Reason: Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could res...
Z-1075-2021February 17, 2021Athens, TX
Arrow International Inc
Class IITerminated
Arrow AGB+ Multi-Lumen CVC Kit 7Fr 3 Lumen- Intended for venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. Product Code: CDC-45703-1A
Reason: The lidstock / information card on the kits correctly states the product code as CDC-45703-1A. The banner card on the side of the kit may incorrectly state the product code as CDC-...
Z-1106-2021February 17, 2021Reading, PA
Prytime Medical Devices, Inc.
Class IITerminated
PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC, STERILE, 1 EA, SYRINGE, 30CC, STERILE 1 EA, INTRODUCER SHEATH KIT, 7 FR W/.035 SHORT GUIDEWIRE...
Reason: There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G needle. If the 21G needle is used to gain vascular access then the guid...
Z-0959-2021February 17, 2021Boerne, TX
Argon Medical Devices, Inc
Class IITerminated
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18C, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/5f straight ca...
Reason: Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could res...
Z-1076-2021February 17, 2021Athens, TX
Argon Medical Devices, Inc
Class IITerminated
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19, 1/19 GA Flexcore Biopsy Needle, 7F Introducer Sheath W/Curved Metal Stiffener, Tissue Removal Swabs, 5F Straight Catheter,...
Reason: Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could res...
Z-1073-2021February 17, 2021Athens, TX
Baxter Healthcare Corporation
Class IITerminated
Baxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K - Product Usage: intended for single patient u...
Reason: There were customer reports of separation between the tubing and Male Luer Lock Adaptor.
Z-1102-2021February 17, 2021Deerfield, IL
ICU Medical, Inc.
Class IITerminated
154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides access for the adminis...
Reason: Due to the incorrect filter assembly.
Z-0958-2021February 17, 2021San Clemente, CA
Aomori Olympus Co., Ltd.
Class IITerminated
B7-2LA BALLOON CATHETER 13MM DIA 3.2, model no. B7-2LA - Product Usage: Used with endoscopes for injection of contrast medium or other medical fluid into the biliary or pancreatic tract.
Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Z-0964-2021February 17, 2021Kuroishi, N/A
ICU Medical, Inc.
Class IITerminated
7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides access for the...
Reason: Due to the incorrect filter assembly.
Z-0957-2021February 17, 2021San Clemente, CA
Braxton Medical Corporation
Class IIOngoing
B. Braun Introcan IV Safety Catheters, sterile - Product Usage: It is a single-use device to generate intravascular and tissue access to sample blood, monitor blood pressure, or administer fluids and ...
Reason: Recalled products do not have FDA approval for sale in the United States.
Z-0930-2021February 3, 2021Lutz, FL
Stryker Sustainability Solutions
Class IITerminated
Cardiovascular Reprocessed ViewFlex Xtra ICE Catheter, REF D087031, OM St. Jude Medical, Size 9F (3.2mm), Length 90 cm, Sterile EO, Rx Only, Quantity 1, UDI: 07613327376067
Reason: A subset of temporary intracardiac ultrasound catheters are labeled with the incorrect expiration date. They are mislabeled as having a shelf life of two years, while the actual sh...