Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

3,445 results found for catheter Page 39 of 138
Bard Access Systems Inc.
Class II Terminated
REF 9617405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741035989 - Product Usage: is designed for use when central ...
Reason: Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot ...
Z-1659-2021 May 26, 2021 Salt Lake City, UT
Bard Access Systems Inc.
Class II Terminated
REF 7617408, Groshong NXT ClearVue Catheter, Full Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035255 - Product Usage: is designed for use when central venous catheterization is prescribed.
Reason: Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot ...
Z-1654-2021 May 26, 2021 Salt Lake City, UT
ICU Medical, Inc.
Class II Ongoing
REF CL3028T 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated w/ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros , Bag Hanger, Drop-In Red Cap - Product Usage: provides access for the administr...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1571-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CH2060 60mL Syringe w/Spinning Spiros, Red Cap - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s ne...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1577-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CH3020 40" (102 cm) Appx. 4.8 ml, 20 Drop Admin Set w/In-Line Chamber, 2 Spiros", Hanger - Product Usage: provides access for the administration of fluids from a container to a patient s vascular ...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1579-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CL3960 Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros ; w/ChemoLock; Syringe Transfer Set w/MicroClave, w/ChemoLock Port - Product ...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1574-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF Spinning Spiros Closed Male Luer - Product Usage: provides access for the administration of fluids from a container to a patients vascular system through the administration sets needle or cathe...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1569-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF 011-H2864-HSR 180 cm (71 IN)APPX 11.5ml, PUR AMBR, Spiros, CAMERA, REG. FLUSSO, BCV, Y-Clave - Product Usage: provides access for the administration of fluids from a container to a patient s vascu...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1582-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CH3034 5" (13 cm) Bag Spike Adapter w/Spiros" w/Red Cap, Vented Cap - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the a...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1587-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF 011-H3597 42 cm (16.5 IN) APPX 2.6ml, PUR TRANSFER SET, Clave ADDITIVE PORT, BCV, Spiros - Product Usage: provides access for the administration of fluids from a container to a patient s vascular ...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1583-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CL3538 Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock , Spinning Spiros w/Red Cap - Product Usage: provides access for the administration of fluids from a c...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1572-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF 011-CH3967 Admin Set w/ClaveTM, SpirosTM - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle o...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1581-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF 20124-01 30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger LOT 5142971 - Product Usage: provides access for the administration of fluids from a conta...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1567-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CH3011 30" (76 cm) Appx 3.3 mL, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger - Product Usage: provides access for the administration of fluids from a cont...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1578-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CH3625 5" (13 cm) Appx 0.50 ml, Smallbore Trifuse Ext Set w/3 ChemoClave (Purple, Green Rings), Spiros w/Red Cap, 3 Clamps - Product Usage: provides access for the administration of fluids from a...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1580-2021 May 19, 2021 San Clemente, CA
Medtronic Inc.
Class III Ongoing
VenaSeal Closure System
Reason: Firm received increased number of reports that VenaSeal dispensing systems are not effectively advancing the adhesive forward through the delivery catheter.
Z-1564-2021 May 19, 2021 Plymouth, MN
ICU Medical, Inc.
Class II Ongoing
REF CL4151 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer - Product Usage: provides access for the administratio...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1575-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CL3946 Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock , Spinning Spiros w/Red Cap - Product Usage: provides access for the administration of fluids from a c...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1573-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CL3011 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated w/ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger - Product Usage: provides access for the administration of fluids from a container...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1570-2021 May 19, 2021 San Clemente, CA
Bard Peripheral Vascular Inc
Class II Terminated
PowerPort duo M.R.I. Implantable Port, with attachable 9.5F Polurethane Open-Ended Dual-Lumen Venous Catheter, REF 5829502, UDI: (01)00801741027406 - Product Usage: The port system can be used for inf...
Reason: Catheters may experience difficulty in flushing, infusion, and/or aspiration, and septum dislodgements.
Z-1552-2021 May 19, 2021 Tempe, AZ
Bard Peripheral Vascular Inc
Class II Terminated
PowerPort duo M.R.I. Implantable Port, with attachable 9.5F Polurethane Open-Ended Dual-Lumen Venous Catheter, REF 1829500, UDI: (01) 00801741027185 - Product Usage: The port system can be used for in...
Reason: Catheters may experience difficulty in flushing, infusion, and/or aspiration, and septum dislodgements.
Z-1550-2021 May 19, 2021 Tempe, AZ
ICU Medical, Inc.
Class II Ongoing
REF 034-H2629 Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros" - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through th...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1584-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF K7040-001 Spinning Spiros Closed Male Luer - Product Usage: provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1590-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF CH3762 35" (89 cm) Appx 3.9 ml, 10 Drop Admin Set w/Integrated Clave" Drip Chamber, Clave", Spiros" w/Red Cap, Bag Hanger - Product Usage: provides access for the administration of fluids from a ...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1589-2021 May 19, 2021 San Clemente, CA
ICU Medical, Inc.
Class II Ongoing
REF 20130-01 Spinning Spiros¿ Closed Male Luer, Red Cap REF CH2000S-PC Spinning Spiros¿ Closed Male Luer, Purple Cap - Product Usage: provides access for the administration of fluids from a con...
Reason: Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Z-1586-2021 May 19, 2021 San Clemente, CA