Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

2,985 results found for software Page 38 of 120
Philips Ultrasound, Inc.
Class II Completed
Philips Affiniti 70, Diagnostic Ultrasound System Model # 795210
Reason: Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.
Z-0284-2024 November 15, 2023 Bothell, WA
Philips Ultrasound, Inc.
Class II Completed
Philips EPIQ Diagnostic Ultrasound System Models: EPIQ 7G(model # 795200), EPIQ 7C (Model # 795201), EPIQ 7W (Model # 795202), EPIQ 5G (model # 795204), EPIQ 5W (model # 795206), EPIQ CVx (Model # 79...
Reason: Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.
Z-0283-2024 November 15, 2023 Bothell, WA
Asensus Surgical Inc
Class I Terminated
Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Nu...
Reason: Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparos...
Z-0170-2024 November 8, 2023 Durham, NC
Siemens Medical Solutions USA, Inc
Class II Ongoing
***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgical environment and provides comprehensive image acquisition modes. Material Number: 11108110...
Reason: Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patien...
Z-0196-2024 November 8, 2023 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Ongoing
Siemens Cios Alpha (VA30) mobile fluoroscopic C-arm X-ray System, designed for the surgical environment. 11105200
Reason: Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patien...
Z-0195-2024 November 8, 2023 Malvern, PA
GE Healthcare
Class II Ongoing
Centricity Universal Viewer, software versions 6.0 SP0 through 6.0 SP7.1; Radiological Image Processing System
Reason: GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW...
Z-0140-2024 November 1, 2023 Chicago, IL
Acclarent, Inc.
Class II Ongoing
TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z and TDC0005
Reason: When using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation systems intended for use during su...
Z-0127-2024 November 1, 2023 Irvine, CA
Impedimed Limited
Class II Ongoing
SOZO Bilateral Arm L-Dex Software
Reason: Bilateral L-Dex assessment software does not have the same level of sensitivity to help detect early signs of lymphedema as the unilateral arm L-Dex assessment, which could result ...
Z-0155-2024 November 1, 2023 Pinkenba, N/A
GE Healthcare
Class II Ongoing
Centricity PACS-IW, software versions V3.7.x through 3.7.3.9 SP3; Radiological Image Processing System
Reason: GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW...
Z-0138-2024 November 1, 2023 Chicago, IL
OLEA MEDICAL
Class II Ongoing
Functional MR V1.0
Reason: When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting multiple volumes of interest, a functional magnetic resonance software bug may cause ...
Z-0122-2024 October 25, 2023 LA CIOTAT, N/A
EITAN MEDICAL LTD
Class I Ongoing
Sapphire Infusion Pumps, Models: Multi-Therapy (REF 15031-000-0028), Epidural (REF 15032-000-0027) and SapphirePlus (REF 15038-000-0001)
Reason: Infusion Pumps with affected software revision may fail to detect air in line, which may lead to air embolism.
Z-0094-2024 October 25, 2023 Netanya, N/A
Smiths Medical ASD Inc.
Class II Ongoing
Medfusion Syringe Pump, Model 4000-XXXX-XX: a) 0100-50, b) 0100-249, c) 0101-50, d) 0101-51, e) 0101-78, f) 0101-249, g) 0105-51, h) 0105-78, i) 0105-249, j) 0106-00, k) 0106-01, ...
Reason: A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the forc...
Z-0079-2024 October 18, 2023 Minneapolis, MN
RoyalVibe Health Ltd.
Class II Ongoing
CellQuicken Analyzer (Smart-Watch and Software), RoyalVibe Ultrasound Generator, Brainvibe, Envirovibe Water Restructuring Pad, RoyalVibe Therapy Balls
Reason: Ultrasound devices were not authorized, cleared, or approved by FDA, and thus their performance characteristics have not been adequately established. There is a possibility of cust...
Z-0082-2024 October 18, 2023 The Woodlands, TX
Smiths Medical ASD Inc.
Class II Ongoing
Medfusion Syringe Pump, Model 3500-0600-XX: a) 0600-00; b) 0600-01; c) 0600-50; d) 0600-51; e) 0600-82; f) 0600-249; software versions: v3, v4, v5, and v6
Reason: A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the forc...
Z-0078-2024 October 18, 2023 Minneapolis, MN
Mazor Robotics Ltd
Class II Ongoing
MAZOR X robotic guidance system, Model TPL0059, with software versions 4.2.2 or 5.0.1
Reason: Software update
Z-0101-2024 October 18, 2023 Caesarea, N/A
Blue Belt Technologies, Inc
Class II Ongoing
Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
Reason: CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon choo...
Z-0041-2024 October 11, 2023 Pittsburgh, PA
Medtronic Neuromodulation
Class II Terminated
Restore Clinician Programmer Application Software, model number A71100, used with the Restore Neurostimulators.
Reason: The original version of the A71100 Restore Clinician Programmer Application has been identified to have a compatibility issue with some legacy clinician programmer software resulti...
Z-2657-2023 October 11, 2023 Minneapolis, MN
Philips North America Llc
Class II Ongoing
EarlyVue VS30 Vitals Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863380
Reason: A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.
Z-2654-2023 October 11, 2023 Cambridge, MA
Philips North America Llc
Class II Ongoing
EarlyVue VS30 Vital Signs Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863359
Reason: A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.
Z-2655-2023 October 11, 2023 Cambridge, MA
Philips Healthcare
Class II Ongoing
Incisive CT, software version 5.0
Reason: Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the opti...
Z-2520-2023 October 11, 2023 Cambridge, MA
Fresenius Kabi USA, LLC
Class II Ongoing
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled ad...
Reason: Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment w...
Z-2635-2023 October 4, 2023 North Andover, MA
CareFusion 303, Inc.
Class II Ongoing
BD Pyxis MedBank MedPass Software, REF: 139088-01
Reason: After dispensing a timed dose medication for a patient, the automated dispensing cabinet software may continue to show the already dispensed medication dose is still available for ...
Z-2646-2023 October 4, 2023 San Diego, CA
Baxter Healthcare Corporation
Class II Ongoing
Baxter Epiphany Cardio Server E3 ECG Management System Servers with software versions: a) v6.1.x, b) v6.2.x, c) v7.0.x.
Reason: Epiphany, now a Baxter Healthcare company, is issuing an Urgent Medical Device Correction for the Cardio Server E3 ECG Management System software listed below due to a report of th...
Z-2647-2023 October 4, 2023 Deerfield, IL
Exocad GmbH
Class II Ongoing
exoplan version 3.1 Rijeka Software -A medical software, intended to support the pre-operative planning of dental implants using the visualization of the implant placement within images of the patents...
Reason: A software library filtering error has been discovered which occurs when users are using the Step-by-Step Full Drill Protocol export functionality of exoplan 3.1 Rijeka. When a gui...
Z-2633-2023 October 4, 2023 Darmstadt, N/A
Philips North America Llc
Class II Ongoing
Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number...
Reason: Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardw...
Z-2608-2023 September 27, 2023 Cambridge, MA