Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

2,985 results found for software Page 35 of 120
Electro Medical Systems SA
Class II Ongoing
Brand Name: PIEZON Bottle Product Name: PIEZON Bottle Model/Catalog Number: EG-111 Software Version: N/A Product Description: These bottles are sold empty and are filled by the user with water - i...
Reason: Received complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or storage if the use is not...
Z-1688-2024 May 1, 2024 Nyon, N/A
Siemens Medical Solutions USA, Inc.
Class II Ongoing
ACUSON Juniper Diagnostic Ultrasound System, REF: 11335791, and ACUSON Juniper Select Diagnostic Ultrasound System, REF: 11653093, running software version VA10x with DICOM SR Cardiac option, from VB1...
Reason: On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and resul...
Z-1599-2024 May 1, 2024 Issaquah, WA
SonarMed Inc
Class I Ongoing
AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.
Reason: Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.
Z-1535-2024 April 24, 2024 Carmel, IN
Philips Respironics, Inc.
Class I Ongoing
Trilogy Evo Universal Ventilator, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
Reason: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Z-1505-2024 April 24, 2024 Murrysville, PA
Philips Respironics, Inc.
Class I Ongoing
Trilogy Evo O2, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
Reason: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Z-1503-2024 April 24, 2024 Murrysville, PA
Philips Respironics, Inc.
Class I Ongoing
Trilogy EV300, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
Reason: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Z-1502-2024 April 24, 2024 Murrysville, PA
Philips Respironics, Inc.
Class I Ongoing
Trilogy Evo, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
Reason: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Z-1504-2024 April 24, 2024 Murrysville, PA
Philips Respironics, Inc.
Class I Ongoing
Garbin EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)
Reason: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Z-1500-2024 April 24, 2024 Murrysville, PA
Philips Respironics, Inc.
Class I Ongoing
Aeris EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)
Reason: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Z-1499-2024 April 24, 2024 Murrysville, PA
Philips Respironics, Inc.
Class I Ongoing
LifeVentEVO2, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)
Reason: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Z-1501-2024 April 24, 2024 Murrysville, PA
Philips North America
Class II Ongoing
PHILIPS Incisive Computed tomography X-ray system Incisive CT Software Upgrade from V4.5.1.41921 to V4.5.1.41989-Intended to produce cross-sectional images of the body Model Number: 728143
Reason: Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be ...
Z-1507-2024 April 17, 2024 Cambridge, MA
RAYSEARCH LABORATORIES AB
Class II Ongoing
RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
Reason: A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
Z-1482-2024 April 17, 2024 Stockholm, N/A
Fresenius Kabi USA, LLC
Class I Ongoing
LVP Software of the Ivenix Infusion System (IIS), Version 5.8.0.
Reason: Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.
Z-1484-2024 April 17, 2024 North Andover, MA
Life Technologies Corporation
Class II Ongoing
Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and ...
Reason: Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems exploitation of the vulnerability by a threat actor may allow them to alter settings,...
Z-1426-2024 April 10, 2024 Frederick, MD
Philips North America
Class II Ongoing
CareEvent Event Management System, Model Numbers 866435 and 866436 (CareEvent Upgrade), Software Version Number C.03.X
Reason: It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.
Z-1387-2024 April 3, 2024 Cambridge, MA
Philips North America
Class II Ongoing
Patient Information Center iX (PIC iX), Model Numbers 866389, 866390 (PIC iX Expand), and 866424 (PIC iX Hardware) running Software Version Number 4.X
Reason: It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.
Z-1386-2024 April 3, 2024 Cambridge, MA
Landauer
Class II Completed
microStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110, and 201), 10016-000 (marketed in EMEA as 101), 10036-000, 10044-000, 10055-000, BC30058, and BC30059
Reason: Permanent discontinuance of the microSTARii medical dosimetry system. Landauer has initiated this action to remove all reference to use of the microStar readers with the nanoDot o...
Z-1349-2024 March 27, 2024 Glenwood, IL
Fresenius Kabi USA, LLC
Class II Completed
LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
Reason: Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, res...
Z-1282-2024 March 20, 2024 North Andover, MA
Fresenius Kabi USA, LLC
Class II Completed
LVP software of the Ivenix Infusion System (IIS); LVP-SW-0004
Reason: Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is execut...
Z-1283-2024 March 20, 2024 North Andover, MA
Ortho-Clinical Diagnostics, Inc.
Class II Ongoing
VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)
Reason: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause ...
Z-1226-2024 March 13, 2024 Rochester, NY
Ortho-Clinical Diagnostics, Inc.
Class II Ongoing
VITROS XT 7600 Integrated System, Catalog No. 6844461
Reason: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause e...
Z-1228-2024 March 13, 2024 Rochester, NY
Ortho-Clinical Diagnostics, Inc.
Class II Ongoing
VITROS XT 3400 Chemistry System, Catalog No. 6844458
Reason: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is...
Z-1232-2024 March 13, 2024 Rochester, NY
Ortho-Clinical Diagnostics, Inc.
Class II Ongoing
VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished)
Reason: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause ...
Z-1224-2024 March 13, 2024 Rochester, NY
Ortho-Clinical Diagnostics, Inc.
Class II Ongoing
VITROS XT 3400 Chemistry System, Catalog No. 6844458
Reason: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause e...
Z-1227-2024 March 13, 2024 Rochester, NY
Ortho-Clinical Diagnostics, Inc.
Class II Ongoing
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Reason: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause e...
Z-1225-2024 March 13, 2024 Rochester, NY