Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IIOngoing
CRT-P Quad Models (CRT-Ps): a. Percepta CRT-P MRI: W1TR01, W1TR04, W4TR01, W4TR04 b. Serena CRT-P MRI: W1TR02, W1TR05, W4TR02, W4TR05 c. Solara CRT-P MRI: W1TR03, W1TR06, W4TR03, W4TR06. implan...
Reason: There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device tri...
Z-1523-2021May 12, 2021Mounds View, MN
Coltene Whaledent Inc
Class IIOngoing
Dental pins - TMS LINK PLUS Self-Threading Pins - Product Usage: implanted into the teeth or for lab use for dental restoration purposes. The normal procedure requires a hole to be drilled into the ti...
Reason: The incorrect dental pins were packaged and subsequently distributed.
Z-1494-2021May 5, 2021Cuyahoga Falls, OH
Omnia SRL
Class IITerminated
Omnia Soft Tissue Punch is a dental instrument designed to remove with precision a circular section of oral soft tissue (e.g., gum, palatal mucosa) during dental surgery, typically to allow access to ...
Reason: Omnia SRL received a notification from the contract manufacturer informing them of inadequate sterilization processes which have been taking place at their contracted sterilizing f...
Z-1518-2021May 5, 2021Fidenza, N/A
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ILIVIA NEO 7 VR-T DF-1 ProMRI, REF 429531, UDI: 04035479156763 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of l...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1462-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK Iforia 7,DR-T, DF-1,REF 390064, UDI: 04035479126926 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threate...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1448-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ILIVIA NEO 7 DR-T DF-1 ProMRI, REF 429529, UDI: 04035479156749 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of l...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1464-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK IPERIA 7, DR-T DF4 ProMRI, REF 392423, UDI: 04035479128913 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1453-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK Ilestro 7 DR-T, DF-4, REF383564, UDI: 04035479125202 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threat...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1442-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK Ilestro 7 VR-T, DF-1, REF 383578, UDI: 04035479125349 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threa...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1443-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ILIVIA 7, DR-T DF4 ProMRI, REF 404623, UDI: 04035479142100 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1456-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK INTICA NEO 7 VR-T DX DF-1 ProMRI, REF 429559, UDI: 04035479156855 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment o...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1463-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK Iforia 7, vr-T, DX, DF-1,REF 390090, UDI: 04035479127176 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-th...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1447-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ITREVIA 7, VR-T DX DF-1, REF 393037, UDI: 04035479129521 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-th...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1450-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ILIVIA 7, DR-T DF-1 PROMRI, REF 404622, UDI: 04035479142094 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1461-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ILIVIA 7, VR-T DF4 ProMRI, REF 404626, UDI: 04035479142131 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1458-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ITREVIA 7, DR-T QP DF-1, REF 392412, UDI: 04035479128807 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-t...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1455-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK IPERIA 7,VR-T, DX, DF-1, ProMRI, REF 393032, UDI: 04035479129477 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1449-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK INVENTRA 7, VR-T DX DF-1 ProMRI, REF 399436, UDI: 04035479136314 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment o...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1459-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK Ilestro 7 DR-T, DF-1, REF383562, UDI: 04035479125189 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threat...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1441-2021April 28, 2021Lake Oswego, OR
Heartware, Inc.
Class IOngoing
Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these system components: HVAD Sterile Implant Kit, Model Numbers: 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, MCS1705PU...
Reason: Medtronic HeartWare HVAD System Instructions for Use (IFU) and Patient Manual (PM) updates regarding: Carrying Cases, Driveline Cover Orientation, and Controller Power Up Sequence ...
Z-1426-2021April 28, 2021Miami Lakes, FL
BIOTRONIK Inc
Class IITerminated
BIOTRONIK INVENTRA 7, HF-T QP DF4 IS4, REF 393012, UDI: 04035479129286 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of li...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1454-2021April 28, 2021Lake Oswego, OR
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ITREVIA 7, VR-T DF-4, REF 393041, UDI: 04035479129569 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threa...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1452-2021April 28, 2021Lake Oswego, OR
Heartware, Inc.
Class IIOngoing
Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Hear...
Reason: IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very shor...
Z-1469-2021April 28, 2021Miami Lakes, FL
BIOTRONIK Inc
Class IITerminated
BIOTRONIK ILIVIA 7, VR-T DF-1 ProMRI, REF 404625, UDI: 04035479142124 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of lif...
Reason: There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Z-1457-2021April 28, 2021Lake Oswego, OR
Heartware, Inc.
Class IIOngoing
Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. ...
Reason: IFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connecto...