Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1791-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-DR DDME3D1 MIRRO MRI, Model Number DDME3D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1719-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D DTBB2D1 VIVA S IS1/DF1 INTL, Model Number DTBB2D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1745-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D VIVA QUAD XT GOLD CTD, Model Number DTBA2QQG; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1740-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D DTBA1QQ VIVA QUAD XT IS4/DF4 US, Model Number DTBA1QQ; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1733-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-VR DVAB1D1 VISIA AF US IS1/DF1, Model Number DVAB1D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1789-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D DTBB1D1 VIVA S IS1/DF1 US, Model Number DTBB1D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1741-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-VR DVAB1D4 VISIA AF US DF4, Model Number DVAB1D4; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1790-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D DTBA1Q1 VIVA QUAD XT IS4/DF1 US, Model Number DTBA1Q1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1732-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRTD DTMA2QQ CLARIA MRI QUAD OUS DF4, Model Number DTMA2QQ; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1760-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-DR DDME3D4 MIRRO MRI, Model Number DDME3D4; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1720-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD CROME VR MRI DF4, Model Number DVPC3D4; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1823-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRTD DTMC1D1 COMPIA MRI US DF1, Model Number DTMC1D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1769-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRTD DTMB2D1 AMPLIA MRI OUS DF1, Model Number DTMB2D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1765-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-VR DVBC3D1 EVERA S IS1/DF1 GLOB, Model Number DVBC3D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1799-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRTD DTMB1QQ AMPLIA MRI QUAD US DF4, Model Number DTMB1QQ; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1764-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRTD COBALT XT HF QUAD MRI IS4 DF4, Model Number DTPA2QQ; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1779-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-VR DVBB2D1 EVERA XT IS1/DF1 INTL, Model Number DVBB2D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1797-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-VR DVBB1D1 EVERA XT IS1/DF1 US, Model Number DVBB1D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1795-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD COBALT DR MRI IS1 DF1, Model Number DDPB3D1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1725-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D DTBB1Q1 VIVA QUAD S IS4/DF1 US, Model Number DTBB1Q1; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1743-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRTD DTMB2QQ AMPLIA MRI QUAD OUS DF4, Model Number DTMB2QQ; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1768-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD COBALT XT HF QUAD OUS, Model Number DTPA2QQG; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1780-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRTD DTPA2D1G COBALT XT HF QUAD OUS, Model Number DTPA2D1G; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1775-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D DTBB2QQ VIVA QUAD S IS4/DF4 INTL, Model Number DTBB2QQ; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...