Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

3,445 results found for catheter Page 20 of 138
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO IMT (5PK), REF H749085262022; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1650-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO CLS4 (5PK), REF H74908526575A2; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1679-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO MULTIPACK 145 (5PKX3), REF H749085265022; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1673-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL4 (5PK), REF H74908526222; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1655-2024 May 22, 2024 Maple Grove, MN
Sophysa S.A.
Class II Ongoing
The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventric...
Reason: An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could p...
Z-1825-2024 May 22, 2024 Orsay, N/A
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL4.5 (5PK), REF H74908526232; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1658-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR5 (SGL), REF H74908526031; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1640-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO JIG4 (5PK), REF H749085262532; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1664-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO AR2 (5PK), REF H74908526862; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1681-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO AR MOD (5PK), REF H74908526892; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1682-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FR5 (5PK), REF H74908526032; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1641-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO MULTIPACK 155 (5PKX3), REF H749085263022; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1669-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO WR (SGL), REF H74908526501; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1671-2024 May 22, 2024 Maple Grove, MN
Boston Scientific Corporation
Class I Ongoing
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO CLS2.5 (5PK), REF H74908526572A2; cardiac catheter
Reason: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delamina...
Z-1676-2024 May 22, 2024 Maple Grove, MN
Stryker Neurovascular
Class II Ongoing
The Distal Access Catheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophil...
Reason: Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was re...
Z-1740-2024 May 8, 2024 Fremont, CA
ROUTE 92 MEDICAL INC
Class I Ongoing
Route 92 Medical 070 Reperfusion System, REF 1003, Rxonly, SterileEO. Used for introduction of interventional devices into the neurovasculature and aspiration of thrombus in ischemic stroke patients.
Reason: Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
Z-1614-2024 May 1, 2024 San Mateo, CA
Covidien LP
Class II Ongoing
Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, apheresis, and infusion...
Reason: Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
Z-1635-2024 May 1, 2024 Mansfield, MA
ROUTE 92 MEDICAL INC
Class I Ongoing
Route 92 Medical, 070 Access System, REF 1004, Rxonly , Sterile EO. Used for introduction of interventional devices into the neurovasculature.
Reason: Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
Z-1612-2024 May 1, 2024 San Mateo, CA
ROUTE 92 MEDICAL INC
Class I Ongoing
Route 92 Medical, Full Length 070 Reperfusion System, REF: 7001, Rxonly, SterileEO. Used for Introduction of interventional devices into the neurovasculature.
Reason: Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
Z-1611-2024 May 1, 2024 San Mateo, CA
Covidien LP
Class II Ongoing
Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, Heparin Coating, Silver Ion Sleeve and Tal VenaTrac Stylet- Intended for acute and chronic hemodialy...
Reason: Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
Z-1634-2024 May 1, 2024 Mansfield, MA
ROUTE 92 MEDICAL INC
Class I Ongoing
Route 92 Medical, Tenzing 7 Delivery Catheter, REF 1005, Rx Only, Sterile EO. Used for introduction of interventional devices into the neurovasculature.
Reason: Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
Z-1613-2024 May 1, 2024 San Mateo, CA
ROUTE 92 MEDICAL INC
Class I Ongoing
Route 92 Medical Full Length 070 Access System, REF 7002, Rxonly, SterileEO. Used for Introduction of interventional devices into the neurovasculature.
Reason: Their is the potential for distal tip separation at the proximal marker band of the delivery catheters.
Z-1610-2024 May 1, 2024 San Mateo, CA
Cook Incorporated
Class II Ongoing
Peel-Away Introducer - Intended for the percutaneous introduction of balloon, electrode and closed or non-tapered end catheters into central and peripheral vasculature, and for nonvascular use. RE...
Reason: Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Z-1528-2024 April 24, 2024 Bloomington, IN
Cook Incorporated
Class II Ongoing
Entuit Secure Adjustable Gastrointestinal Suture Anchor Set - Intended for anchoring the anterior wall of the stomach to the abdominal wall prior to introduction of interventional catheters. REFE...
Reason: Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Z-1529-2024 April 24, 2024 Bloomington, IN
Cook Incorporated
Class II Ongoing
Dilator With AQ Hydrophilic Coating - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the...
Reason: Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Z-1531-2024 April 24, 2024 Bloomington, IN