LifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN, PIGGYBACK, 2 CLAVES PORTS AND OPTION LOK, NON-DEHP; NOMINAL LENGTH: 97 IN (249 CM); PRIMING VOLUME: 14 ML; For the ad...
Reason: Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that have a drip chamber with a ball valve. Preliminary investigation findi...
Z-1751-2013July 24, 2013Lake Forest, IL
Codman & Shurtleff, Inc.
Class ITerminated
MedStream Programmable Infusion Pump , Implantable for the intrathecal delivery of Baclofen or Morphine Product Code: 91-4201US (40ml pump) 91-4201 (40ml pump) Intended for the intrathecal deliv...
Reason: Miscalibrated Fill Level Sensor may affect dosing level
Z-1691-2013July 24, 2013Raynham, MA
Medtronic MiniMed
Class ITerminated
Medtronic MiniMed Paradigm Insulin Infusion Sets The Paradigm infusion sets are intended for use with Paradigm insulin infusion pumps.
Reason: Medtronic is recalling the Medtronic MiniMed Paradigm Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion sets because of a potential safety issue that can occur...
Z-1668-2013July 17, 2013Northridge, CA
Baxter Healthcare Corp.
Class IITerminated
Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication thera...
Reason: There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicate...
Z-1677-2013July 17, 2013Deerfield, IL
Baxter Healthcare Corp.
Class IITerminated
Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication thera...
Reason: There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicate...
Z-1678-2013July 17, 2013Deerfield, IL
Baxter Healthcare Corp.
Class IITerminated
Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication thera...
Reason: There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicate...
Z-1679-2013July 17, 2013Deerfield, IL
Medtronic Neuromodulation
Class ITerminated
Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and deli...
Reason: Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal...
Z-1579-2013July 3, 2013Minneapolis, MN
Medtronic Neuromodulation
Class ITerminated
Medtronic SynchroMed EL Programmable pumps, Models 8626, 8626L, 8627, 8627L, (The SynchroMed EL Pumps are supplied in 10 ml or 18 ml reservoir size.) Sterilized using ethylene oxide. Product Usa...
Reason: Medtronic is providing Healthcare Professionals with important safety information and patient management recommendations regarding the unintended delivery of drug during the primin...
Z-1571-2013July 3, 2013Minneapolis, MN
Medtronic Neuromodulation
Class ITerminated
Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: catheter interface with attached sutureless pump connector, catheter, connector pin, and strain-relief sleeve to be used with ...
Reason: The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the...
Reason: Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the ...
Z-1609-2013July 3, 2013Round Lake, IL
Medtronic Neuromodulation
Class ITerminated
Medtronic SynchroMed EL Programmable pumps, Models 8626, 8626L, 8627, 8627L, (The SynchroMed EL Pumps are supplied in 10 ml or 18 ml reservoir size.) Sterilized using ethylene oxide. The pump is part ...
Reason: Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal...
Z-1580-2013July 3, 2013Minneapolis, MN
Greatbatch Medical
Class IITerminated
Cordis a Johnson & Johnson Company, RadialSource Spring Wire Transradial Access Kit. Manufacturer Greatbatch Medical 2300 Berkshire Lane North, Minneapolis, MN 55441, Distributor Cordis Corporation 14...
Reason: Cordis Corporation, Inc and Greatbatch Medical are recalling one lot of the cordis RADIALSOURCE Spring Wire Transradial Access Kit product. Some units were manufactured with an un...
Z-1608-2013July 3, 2013Minneapolis, MN
Baxter Healthcare Corp.
Class IITerminated
Brand Name: Intermate Infusion Pump. Indicated for the intravenous administration of medications.
Reason: Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product a...
Z-1604-2013July 3, 2013Round Lake, IL
Baxter Healthcare Corp.
Class IITerminated
Brand Name: Intermate Infusion Pump. Indicated for the intravenous administration of medications.
Reason: Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product a...
Z-1603-2013July 3, 2013Round Lake, IL
Medtronic Neuromodulation
Class ITerminated
Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. Product Usage: The pump is part of an infusion sys...
Reason: Medtronic is providing Healthcare Professionals with important safety information and patient management recommendations regarding the unintended delivery of drug during the primin...
Z-1570-2013July 3, 2013Minneapolis, MN
Symbios Medical Products, LLC
Class ITerminated
Part 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump Kit 510201-BP, 5 fenestrated catheter & BIOPATCH; Go Pump Kit 510349-BP, Epidural Catheter, B...
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1454-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part 510121 Disposable Infusion Pump, ...Elastomeric Pump 300 mL, 4mL/hr,...sterile.. found in the following kits GoPump Kit 510203-BP, Epidural catheter & BIOPATCH GoPump Kit 510204, 2.5" Fenest...
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1455-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part510447 Disposable Infusion Pump, ...Elastomeric Pump 600 mL, 2mL/hr/side,... found in the following kits GoPump Kit 510462-BP, 5 " catheter & BIOPATCH GoPump Kit 510558-BP 10 " catheter& BIO...
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1456-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part 510636 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 300 ml perside), 2 to 10 ml/hr. Delivery time 60 to 144 hours; found in the following kits: GoBlock Kit 510646, Disposable Inf...
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1462-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part 510076 Disposable Dual Infusion Pump, ...Elastomeric Dual Pump 300 mL(150 ml/side) 2mL/hr/side,... Sterile, found in the following kits: GoPump kit 510080-BP, 5 fenestrated catheter & BIOPATCH;...
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1453-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part 510613 Disposable Infusion Pump, ...Elastomeric Pump 300mL (150 mL side)( 6mL/hr ),... found in the following kits GoBlock Kit 510612, Disposable Infusion Pump for drug delivery. Symbios G...
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1460-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part 510450 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 10mL/hr ),... found in the following kits GoBlock Kit 510458, Disposable Single Infusion Pump
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1458-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part 510613 Disposable Infusion Pump, ...Elastomeric Pump 300mL (150 mL side)( 6mL/hr ),... found in the following kits GoBlock Kit 510612, Disposable Infusion Pump for drug delivery. Symbios G...
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1461-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part 510490 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 300 ml per side)( 6mL/hr ),... found in the following kits GoBlock Kit 510491, Disposable Single Infusion Pump for drug delivery
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...
Z-1459-2013June 12, 2013Indianapolis, IN
Symbios Medical Products, LLC
Class ITerminated
Part 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kits GoBlock Kit 510472, Disposable Single Infusion Pump. Symbios GOPump Rapid Recovery Syste...
Reason: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured wi...