TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
Reason: A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGR...
Z-2444-2023August 30, 2023Memphis, TN
Boston Scientific Corporation
Class IIOngoing
EMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator
Reason: There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.
Z-2441-2023August 30, 2023Saint Paul, MN
Smith & Nephew, Inc.
Class IIOngoing
TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
Reason: A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGR...
Z-2443-2023August 30, 2023Memphis, TN
Olympus Corporation of the Americas
Class IIOngoing
HF Resection Electrodes - HF resection electrodes are part of a resectoscope system for endoscopic diagnosis and treatment in urological or gynecological applications. 64 Models: 1) A22201C 2) ...
Reason: Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, s...
Z-2413-2023August 23, 2023Center Valley, PA
MEDLINE INDUSTRIES, LP - Northfield
Class IIOngoing
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as STERILE SET FOR IMPLANT-LF, Model Number: OST011F.
Reason: Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
Z-2308-2023August 9, 2023Northfield, IL
MEDLINE INDUSTRIES, LP - Northfield
Class IIOngoing
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ANGIOGRAPHY PACK, Model Number: DYNJ65963A; b. ARTHOGRAM TRAY, Model Number: P...
Reason: Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
Z-2338-2023August 9, 2023Northfield, IL
MEDLINE INDUSTRIES, LP - Northfield
Class IIOngoing
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 18GX6" FEMORAL ART LINE KIT, Model Number: ART960; b. 20Gx3" RADIAL ART LINE KIT, ...
Reason: Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
Reason: Incorrect product code marked on the product, etched with product code MD611, however; the correct product code is MD610.
Z-2146-2023July 26, 2023Center Valley, PA
MEDLINE INDUSTRIES, LP - Northfield
Class IIOngoing
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CHP COCHLEAR SUPPLEMENT CDS, Model Number CDS983069I; b) HEAD AND NECK PACK, Model Number DYNJ30245; c) I...
Reason: Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Z-2172-2023July 26, 2023Northfield, IL
MEDLINE INDUSTRIES, LP - Northfield
Class IIOngoing
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: STERILE SET FOR IMPLANT-LF, Model Number OST011F
Reason: Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Z-2171-2023July 26, 2023Northfield, IL
MEDLINE INDUSTRIES, LP - Northfield
Class IIOngoing
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CRANIOTOMY CDS-1, Model Number CDS982889I; b) JOINT INJECTION TRAY, Model Number DYNDH1722; c) LOWER EXT...
Reason: Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Z-2191-2023July 26, 2023Northfield, IL
Medtronic Sofamor Danek USA Inc
Class IIOngoing
Medtronic Catalyft LS Expandable Interbody System Implant Kit, Anterior Standalone Cage, REF SPS03188, spinal implant kit
Reason: There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted betw...
Reason: There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted betw...
Reason: There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted betw...
Reason: There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted betw...
Reason: There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted betw...
Reason: There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted betw...
Reason: There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted betw...
Z-2119-2023July 19, 2023Memphis, TN
T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.
Class IIOngoing
BETTA LINK LG KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number: 1...
Reason: Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients include...
Z-2061-2023July 12, 2023GAATON, N/A
T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.
Class IIOngoing
BETTA LINK SR KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number:11...
Reason: Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients include...
Z-2060-2023July 12, 2023GAATON, N/A
Medtronic Neuromodulation
Class IIOngoing
Vanta with AdaptiveStim Implantable Neurostimulator, Model 977006; spinal chord stimulation
Reason: A cardioversion procedure may damage the electronics in the Vanta INS device, making the INS unresponsive and non-functional. As of April 19, 2023, Medtronic has received a total o...
Z-2036-2023July 5, 2023Minneapolis, MN
Implant Direct Sybron Manufacturing LLC
Class IIOngoing
LEGACY SMARTBASE, NE 3.0D 1MML, REF: 8730-81NE; LEGACY SMARTBASE, NE 3.0D 2MML, REF 8730-82NE; LEGACY SMARTBASE NARROW NE 3.5D, REF: 8735-80NNE; LEGACY SMARTBASE, NE 3.5D 2MML, REF: 8735-82NE LEGA...
Reason: Non-engaging abutments were distributed without FDA clearance; the performance characteristics have not yet been adequately established.
Z-2037-2023July 5, 2023Westlake Village, CA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
ICD-VR DVFB1D4 VISIA MRI AF US DF4, Model Number DVFB1D4; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1804-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D DTBB2D4 VIVA S IS1/DF4 INTL, Model Number DTBB2D4; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...
Z-1746-2023June 28, 2023Mounds View, MN
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class IOngoing
CRT-D DTBB1D4 VIVA S IS1/DF4 US, Model Number DTBB1D4; Implantable Cardioverter Defibrillators
Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchr...