Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

3,445 results found for catheter Page 17 of 138
Merit Medical Systems, Inc.
Class II Ongoing
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: ReSolve (Catheter and Tip Suction kit), REF: RTT14038MB/A The ReSolve Thoracosto...
Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that ma...
Z-2570-2024 August 21, 2024 South Jordan, UT
Argon Medical Devices, Inc
Class II Ongoing
L-CATH PICC S/L Peripherally Inserted Central Catheter, 26Ga (1.9F) 0.60mm x 30cm, REF: 384539
Reason: Peripherally Inserted Central Catheter (PICC) is labeled 26G/1.9Fr x 30cm but package contains incorrect 26G/1.9Fr x 50cm PICC, which could lead to procedural delay.
Z-2638-2024 August 21, 2024 Athens, TX
Merit Medical Systems, Inc.
Class II Ongoing
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: SafeSheath CSG/Worley advanced CSG (Introducer Catheter kit), REF: FCL-083-02/A W...
Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that ma...
Z-2577-2024 August 21, 2024 South Jordan, UT
Merit Medical Systems, Inc.
Class II Ongoing
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Outake, REF CET200 and CET300, The Catheter Extractor is a catheter accessory d...
Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that ma...
Z-2559-2024 August 21, 2024 South Jordan, UT
Merit Medical Systems, Inc.
Class II Ongoing
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF:K04-00791A An intravascular administration set is ...
Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that ma...
Z-2561-2024 August 21, 2024 South Jordan, UT
Merit Medical Systems, Inc.
Class II Ongoing
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: SafeSheath CSG (Introducer Catheter), REF: FCL-069-00/A FCL-069-02/A FCL-069-03/...
Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that ma...
Z-2571-2024 August 21, 2024 South Jordan, UT
Merit Medical Systems, Inc.
Class II Ongoing
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude Prestige (Introducer Catheter), REF: PLP-2505/A PLP-2514/A PLP-2516/A ...
Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that ma...
Z-2572-2024 August 21, 2024 South Jordan, UT
Merit Medical Systems, Inc.
Class II Ongoing
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K08-PI (Custom Fluid Management Kit), REF: K08-02042 K08-02948 K08-03026 K08-03...
Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that ma...
Z-2563-2024 August 21, 2024 South Jordan, UT
MEDLINE INDUSTRIES, LP - Northfield
Class I Ongoing
Medline procedure kits labeled as: 1) WMC HYBRID OR PACK PART 1 , Pack Number CMPJ08068F ; 2) VASCULAR PACK , Pack Number CMPJ11455B ; 3) SALEM CATH LAB-LF , Pack Number DYNJ26938J ; 4) EH PE...
Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that ma...
Z-2444-2024 August 14, 2024 Northfield, IL
Angiodynamics, Inc.
Class II Ongoing
SOFT-VU KUMPE 5F X 40CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732702
Reason: Mislabeled. The shelf-box labeling (unit of sale) is accurate for the packaged devices, however the inner pouch label may contain information for a different model (catheter length...
Z-2453-2024 August 7, 2024 Queensbury, NY
Angiodynamics, Inc.
Class II Ongoing
SOFT-VU KUMPE 5F X 65CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732701
Reason: Mislabeled. The shelf-box labeling (unit of sale) is accurate for the packaged devices, however the inner pouch label may contain information for a different model (catheter length...
Z-2454-2024 August 7, 2024 Queensbury, NY
Becton Dickinson Infusion Therapy Systems, Inc.
Class II Ongoing
BD Insyte Autoguard BC, Shielded IV Catheter with Blood Control Technology, REF: 382533, 20 GA x 1.00 in (1.1x25 mm) 63 mL/min, Rx Only. BD Insyte Autoguard BC Shielded IV Catheter is intended to be i...
Reason: Catheters may contain a hole in the catheter tubing, which could result in leakage during the insertion process.
Z-2458-2024 August 7, 2024 Sandy, UT
Siemens Medical Solutions USA, Inc
Class II Ongoing
Sensis Vibe systems with software version VD15B in combination with Sensis High End Server -Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or e...
Reason: If the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possib...
Z-2438-2024 July 31, 2024 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Ongoing
Sensis Vibe systems with software version VD15B-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as wel...
Reason: If the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possib...
Z-2441-2024 July 31, 2024 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Ongoing
Sensis Vibe systems with software version VD15B in combination with VM VIRTUAL SERVER Model: 10765502-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizati...
Reason: If the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possib...
Z-2439-2024 July 31, 2024 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Ongoing
Sensis Vibe systems with software version VD15B in combination with AXIOM Sensis Post Processing Workstation-Intended to be used as a diagnostic and administrative tool supporting hemodynamic cathet...
Reason: If the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possib...
Z-2440-2024 July 31, 2024 Malvern, PA
Medtronic Neuromodulation
Class II Ongoing
Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
Reason: Design update to the Ascenda Intrathecal Catheter Models 8780, 8781, and 8784. The intent of the design update is to reduce the potential for tissue growth into the Ascenda cathe...
Z-2352-2024 July 24, 2024 Minneapolis, MN
Cardinal Health 200, LLC
Class II Ongoing
Presource kits and trays, labeled as: 1) Cardiac Catheterization Kit, Catalog Number SAN69CCKRB; 2) Cardiac Catheterization Kit, Catalog Number SAN73CCV10; 3) Cardiac Catheterization Kit, Catal...
Reason: Cardinal Health is following the FDA's recommendations regarding the disposition of affected Jiangsu Shenli Medical Production Co. Ltd (Shenli) Syringes in the Merit Medical¿ supp...
Z-2328-2024 July 17, 2024 Waukegan, IL
Angiodynamics, Inc.
Class II Ongoing
IsoLoc Prostate Immobilization Treatment Balloon Device Kit, Catalog/UPN: ISO-100
Reason: A single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch ide...
Z-2267-2024 July 10, 2024 Queensbury, NY
MEDLINE INDUSTRIES, LP - Northfield
Class II Ongoing
Sterile surgical convenience kits: a. Medline Cath Lab Special Procedure, Pack #00-RDF0360, containing a Pure Pouch with component Medline Reorder #DYNJ04048, Kelly Forceps/STD STR Sterile. b....
Reason: Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.
Z-2224-2024 July 10, 2024 Northfield, IL
Stryker Sustainability Solutions
Class II Ongoing
Stryker Sustainability Solutions, BW Lasso 2515 NAV eco Variable Diagnostic EP Catheter, REF: D134301, ELECTRODES: 22, spacing:2-6-2MM, Length 115 cm, STERILE.EO, Rx Only
Reason: Diagnostic electrophysiology (EP) catheters do not meet testing requirements.
Z-2268-2024 July 10, 2024 Phoenix, AZ
Cook Incorporated
Class II Ongoing
Ultrathane Cook-Cope Type Locking Loop Multipurpose Drainage Catheter-ULT12.0-38-45-P-6S-CLM-RH-ntended for percutaneous drainage in a variety of drainage applications (e.g., nephrostomy, biliary and ...
Reason: Product labels state the incorrect expiration dates that extends beyond their true expiration date
Z-2223-2024 July 3, 2024 Bloomington, IN
Mermaid Medical A/S
Class II Ongoing
The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque sp...
Reason: Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.
Z-2213-2024 July 3, 2024 Stenlose
Cook Incorporated
Class II Ongoing
Ultrathane Mac-Loc Locking Loop Multipurpose Drainage Catheter ULT16.0-38-25-P-6S-MCL-HC--Intended for percutaneous drainage in a variety of drainage applications (e.g., nephrostomy, biliary and absce...
Reason: Product labels state the incorrect expiration dates that extends beyond their true expiration date
Z-2222-2024 July 3, 2024 Bloomington, IN
C.R. Bard Inc
Class II Ongoing
Bard Medical SureStep Foley Tray System, 16 Fr., Catalog #A319416AM; and Bard Medical Bardex IC Complete Care Infection Control Temperature-Sensing Foley Catheter Tray with Bard Hydrogel and Bacti-Gua...
Reason: Some of the devices may function incorrectly resulting in no temperature display or erratic temperatures being displayed on the temperature monitoring devices.
Z-2154-2024 June 26, 2024 Covington, GA