Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

2,985 results found for software Page 112 of 120
Del Mar Reynolds Medical, Ltd.
Class II Terminated
Spacelabs ARKON Anesthesia Workstation. Spacelabs ARKON Anesthesia Workstation may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing cir...
Reason: Spacelabs ARKON Anesthesia System is recalled due to software defect. The System Status Computer may incorrectly determine that a communication error has occurred. If this situati...
Z-1771-2013 July 31, 2013 Hertford, N/A
Siemens Medical Solutions USA, Inc
Class II Terminated
ARTIS zee and zeego x-ray, angiographic system with Software Artis VC20x/VC21A that have Patch 7 and a 3D and DSA license installed. Angiographic x-ray system
Reason: Siemens issued a customer safety advisory notice and field safety corrective action for a potential issue on Artis systems with Artis VC20x/VC21A that have Patch 7 and a 3D and DSA...
Z-1782-2013 July 31, 2013 Malvern, PA
Stryker Instruments Div. of Stryker Corporation
Class II Terminated
Stryker TPS Universal Driver REF 5100- 99 Rx , The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone r...
Reason: The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect paramet...
Z-1740-2013 July 24, 2013 Portage, MI
Terumo Cardiovascular Systems Corporation
Class II Terminated
CDI Blood Parameter Monitoring System 500 with Venous blood parameter module. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2...
Reason: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update ...
Z-1734-2013 July 24, 2013 Ann Arbor, MI
Terumo Cardiovascular Systems Corporation
Class II Terminated
CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules and Hematocrit/Saturation probe. Intended for use during cardiopulmonary bypass procedures when continuous...
Reason: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update ...
Z-1730-2013 July 24, 2013 Ann Arbor, MI
Terumo Cardiovascular Systems Corporation
Class II Terminated
CDI Blood Parameter Monitoring System 500 with Arterial blood parameter modules and Hematocrit/Saturation probe. Intended for use during cardiopulmonary bypass procedures when continuous monitorin...
Reason: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update ...
Z-1731-2013 July 24, 2013 Ann Arbor, MI
Stryker Instruments Div. of Stryker Corporation
Class II Terminated
RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment...
Reason: The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect paramet...
Z-1742-2013 July 24, 2013 Portage, MI
Terumo Cardiovascular Systems Corporation
Class II Terminated
CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or ven...
Reason: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offs...
Z-1732-2013 July 24, 2013 Ann Arbor, MI
Technidata S.A.
Class III Terminated
TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: Laboratory Information System (L.I.S)
Reason: A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic ...
Z-1743-2013 July 24, 2013 Montbonnot St Martin, N/A
Stryker Instruments Div. of Stryker Corporation
Class II Terminated
CORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System (TPS").The Stryker Total Performance System (TPS) is intended for usein the cutting, drilling...
Reason: The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect paramet...
Z-1741-2013 July 24, 2013 Portage, MI
Terumo Cardiovascular Systems Corporation
Class II Terminated
CDI Blood Parameter Monitoring System 500 with Arterial blood parameter module. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PC...
Reason: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update ...
Z-1733-2013 July 24, 2013 Ann Arbor, MI
Ortho-Clinical Diagnostics
Class II Terminated
ORTHO¿ Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO¿ Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked p...
Reason: There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and...
Z-1651-2013 July 10, 2013 Rochester, NY
Philips Medical Systems (Cleveland) Inc
Class II Terminated
Computed Tomography X-ray systems Product Usage: The Brilliance CT 64, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems that are intended to produce...
Reason: Philips Healthcare received reports from the field that certain Brilliance CT 64, Ingenuity Core, and Ingenuity Core128 systems running software versions 2.6.1 or 3.5.4 exhibited i...
Z-1646-2013 July 10, 2013 Cleveland, OH
GE Healthcare It
Class II Terminated
The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical labora...
Reason: If environment variables MICRO_CORRECT_MSG_TYPE and CORRECT_MSG_TYPE are set to different values, in Microbiology result entry when released results (Final Report released) are c...
Z-1585-2013 July 3, 2013 Barrington, IL
GE Healthcare, LLC
Class II Terminated
GE Healthcare, PET VCAR. PET VCAR application software is affected if installed on any of the following versions of the Advantage Workstation and AW Server: Advantage Workstation 4.6, Volume Share ...
Reason: GE Healthcare has recently become aware of a potential safety issue with respect to the automated PERCIST response criteria calculation applied in the PET VCAR application on Advan...
Z-1551-2013 June 26, 2013 Waukesha, WI
Hitachi Medical Systems America Inc
Class II Terminated
Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with phys...
Reason: Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error c...
Z-1559-2013 June 26, 2013 Twinsburg, OH
Respironics California Inc
Class I Terminated
Respironics V60 Ventilator, Model #V60 Respironics Material P/N (Philips 12 Digit P/N): 85008 (85008) 1053613 (989805628251) 1053614 (989805612101) 1053615 (989805613391) 1053616 (989805613...
Reason: Respironics California, Inc. has initiated a recall on the V60 ventilator because of an issue with the software on the V60 Power Management Board Assembly. If a component fails on...
Z-1537-2013 June 26, 2013 Carlsbad, CA
Tecan US, Inc.
Class II Terminated
Tecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run previously defined scripts and processes on a Freedom EVO platform (Touch Tools Suite is an...
Reason: WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt...
Z-1560-2013 June 26, 2013 Morrisville, NC
Optovue, Inc.
Class III Terminated
Optovue iVue with Normative Database (NDB) with Software Version 3.0 and 3.1. Intended for in vivo imaging.
Reason: Colors from the NDB comparison for Ganglion Cell Complex thickness in the singular OU report in iVue version 3.0 (US) and 3.1(international) are different from those in the individ...
Z-1441-2013 June 12, 2013 Fremont, CA
American Optisurgical Inc
Class II Terminated
American Optisurgical's TX1 Tissue Removal System Console. The console is a software unit that is indicated for use in surgical procedures.
Reason: The recall has been initiated due to an overheating protection voltage suppressor on the redundant safety protection circuit board. Secondary to the recall, American Optisurgical, ...
Z-1516-2013 June 12, 2013 Lake Forest, CA
Alpine Biomed Aps
Class II Terminated
Natus neurology Dantec KEYPOINT.NET 2.20. Catalog/Part Numbers: 9031A070103, 9031A006502, 9033A071103, 9033A072102, 9031A070102, 9033A007402, 9033A071102, 9033A072103, 9033A0701, 9033A0711HA, 9033...
Reason: Alpine Biomed ApS is recalling Natus neurology Dantec KEYPOINT.NET ver. 2.20 Software due to the presence of a software bug.
Z-1407-2013 June 5, 2013 Skovlunde, N/A
RAYSEARCH LABORATORIES AB
Class II Terminated
RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treat...
Reason: Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avo...
Z-1419-2013 June 5, 2013 Stockholm, N/A
Linvatec Corp. dba ConMed Linvatec
Class II Terminated
Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.
Reason: The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key str...
Z-1340-2013 May 29, 2013 Largo, FL
Roche Diagnostics Operations, Inc.
Class II Terminated
Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 module only. Product Usage: Fully automated immunoassay analyzer intended for the in-vitro qua...
Reason: When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific c...
Z-1368-2013 May 29, 2013 Indianapolis, IN
Hologic, Inc.
Class II Terminated
Fluoroscan Mini C-arm InSight, software versions 1.x, 2.x, 3.x The device is a Mini C-arm Fluroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a pati...
Reason: The audible alarm to alert the operator when cumulative irradiation time reaches five minutes is not accurate and the total cumulative irradiation time is not accurate.
Z-1326-2013 May 22, 2013 Bedford, MA