Medical Device Recalls

39,072 Records

FDA medical device enforcement reports

2,985 results found for software Page 107 of 120
Varian Medical Systems, Inc.
Class II Terminated
Plan Parameters workspace within Eclipse Treatment Planning System Version 11 and 13; ARIA Radiation Oncology versions 11 and 13 Acuity, versions 11 and 13. Manufactured by Varian Medical System...
Reason: An anomaly has been traced to a software deficiency within the Plan Parameters workspace. When a dose relevant parameter is changed and "do no clear MUs and Reference Point dose" i...
Z-0552-2014 January 1, 2014 Palo Alto, CA
Roche Diagnostics Operations, Inc.
Class II Terminated
MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of crude sample material (whole blood, serum, blood cells, ...
Reason: This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch wh...
Z-0536-2014 December 25, 2013 Indianapolis, IN
Baxter Corporation Englewood
Class II Terminated
ABACUS TPN Calculation Software, Baxa: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition) ABACUS is a Window...
Reason: Baxter Healthcare Corporation is issuing a voluntary correction for ABACUS Total parenteral Nutrition (TPN) Calculation Software versions 2.0 through 3.0 due to the possibility tha...
Z-0539-2014 December 25, 2013 Englewood, CO
Omnyx Llc
Class II Terminated
Digital Pathology System (DPS) Software version 1.1 The Omnyx" Integrated Digital Pathology system, which consists of a whole slide scanner and associated software applications, is a scanning, rece...
Reason: There is an issue associated with Digital Pathology System (DPS) Software, in which the annotation tools within the DPS version 1.1 Image Viewer will display incorrect measurements...
Z-0469-2014 December 18, 2013 Pittsburgh, PA
Ge Healthcare It
Class II Terminated
The GE Centricity Web; Software Version 3.0x, is intended for use under the direct supervision of a licensed healthcare practitioner. The Centricity Enterprise Web is an image and information distribu...
Reason: GE Healthcare has become aware of an issue associated with rejected images using the WADO protocol within GE Centricity Web that may impact patient safety. The WADO (Web Access to ...
Z-0454-2014 December 11, 2013 Barrington, IL
Siemens Healthcare Diagnostics, Inc.
Class II Terminated
Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1. Diagnostic device intended to determine a variety of analytes i...
Reason: There is the potential for the vial volume of "in use" Vista calibrators, controls, or sample diluent to incorrectly revert to full volume when removed and reloaded. Dimension Vis...
Z-0430-2014 December 11, 2013 Newark, DE
Philips Medical Systems, Inc.
Class II Terminated
Easy Diagnost Eleva Universal Fluoroscopic applications
Reason: The delivered application software was an outdated version to the model of equipment.
Z-0431-2014 December 11, 2013 Andover, MA
Cincinnati Sub-Zero Products Inc
Class II Terminated
Blanketrol III Data Export Software Version 1.10 The Blanketrol III Data Export Software is intended for use in the electronic transfer of medical device data from the Blanketrol III to a computer,...
Reason: During a customer inquiry regarding a beta version of the Bill Data Export Software, Model USB-127, it was discovered that the wrong software version was released to a customer.
Z-0443-2014 December 11, 2013 Cincinnati, OH
Roche Diagnostics Operations, Inc.
Class II Terminated
ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quanti...
Reason: There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possi...
Z-0385-2014 December 4, 2013 Indianapolis, IN
Siemens Medical Solutions USA, Inc
Class II Terminated
(Siemens Rad Fluoro Uro Systems) Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10 Image intensified fluoroscopic x-ray system
Reason: A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in rad...
Z-0374-2014 December 4, 2013 Malvern, PA
Siemens Medical Solutions USA, Inc
Class II Terminated
Siemens RAD Fluoro Uro system (Luminos dRF and Uroskop Omnia with software version VD10A/G Image intensified Fluoroscopic X-ray system
Reason: When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexpo...
Z-0140-2014 December 4, 2013 Malvern, PA
Siemens Healthcare Diagnostics
Class II Terminated
ADVIA 1650 System Software: 1) V4.01, 2) V3.52. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitori...
Reason: Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or ...
Z-0344-2014 November 27, 2013 Tarrytown, NY
Siemens Healthcare Diagnostics
Class II Terminated
ADVIA 1800 System Software V2.01. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.
Reason: Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or ...
Z-0345-2014 November 27, 2013 Tarrytown, NY
GE Healthcare It
Class II Terminated
The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical labora...
Reason: GE Healthcare has become aware of a potential safety issue associated with the use of GE Centricity Laboratory where panel reflex does not occur if DIP.Specimen is blank. Panel ref...
Z-0330-2014 November 27, 2013 Barrington, IL
GE Healthcare, LLC
Class II Terminated
GE Vivid E9 ultrasound system. GE Vivid E9 is a Track 3 diagnostic ultrasound system, which is primarily intended for cardiac imaging and analysis, but which also includes vascular and general radiol...
Reason: GE became aware of a potential safety issue due to a system slow down and system lock up associated with the software of Vivid E9 ultrasound system. During a critical procedure th...
Z-0373-2014 November 27, 2013 Waukesha, WI
Siemens Healthcare Diagnostics
Class II Terminated
ADVIA 1200 System Software: 1) V2.00, 2) V2.01. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitori...
Reason: Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or ...
Z-0343-2014 November 27, 2013 Tarrytown, NY
Philips Medical Systems North America Inc.
Class II Terminated
EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid flow analysis.
Reason: A problem has been detected with Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. Using the International System Date Format may result in calculation errors...
Z-0368-2014 November 27, 2013 Bothell, WA
Siemens Healthcare Diagnostics
Class II Terminated
ADVIA 2400 System Software V4.01. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.
Reason: Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or ...
Z-0346-2014 November 27, 2013 Tarrytown, NY
Carestream Health Inc.
Class II Terminated
Carestream Vue PACS; Virtual PWR Reading Basic SW 8 Pack The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for...
Reason: Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and M...
Z-0216-2014 November 20, 2013 Rochester, NY
Carestream Health Inc.
Class II Terminated
Carestream Vue PACS; Virtual PWR Reading Basic SW StandAlone The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions...
Reason: Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and M...
Z-0220-2014 November 20, 2013 Rochester, NY
Nobel Biocare Usa Llc
Class II Terminated
Osseocare Pro Console & Set Osseocare Pro, Model 1600870-001, 1700470-001, and 1700471-001. Device Name: Osseocare Pro console & Set Osseocare Pro. Model: 1600870-001, 1700470-001, 1700471-001. M...
Reason: The recall was initiated because there is a potential for malfunction of the OsseoCare Pro when used with the OseeoCare Pro iPad application software release 1.2.0.7.
Z-0273-2014 November 20, 2013 Yorba Linda, CA
Carestream Health Inc.
Class II Terminated
Carestream Vue PACS; UPG MX-RAD SVR 50K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for ...
Reason: Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and M...
Z-0259-2014 November 20, 2013 Rochester, NY
Carestream Health Inc.
Class II Terminated
Carestream Vue PACS; Virtual Access Plus SW, Workgroup The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for...
Reason: Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and M...
Z-0212-2014 November 20, 2013 Rochester, NY
Carestream Health Inc.
Class II Terminated
Carestream Vue PACS; Virtual PWR Reading Basic SW, 8 Pack The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions ...
Reason: Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and M...
Z-0214-2014 November 20, 2013 Rochester, NY
Carestream Health Inc.
Class II Terminated
Carestream Vue PACS; UNLIMITED READING LIC < 25K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for h...
Reason: Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and M...
Z-0226-2014 November 20, 2013 Rochester, NY