ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager fo...
Reason: Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the rep...
Z-0987-2015January 21, 2015Newark, DE
Brainlab AG
Class IITerminated
iPlan RT Dose is a stereotactic radiation treatment planning system that is intended for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, an...
Reason: iPlan RT Radiation Treatment Planning Software: Potentially incorrect patient positioning when using multiple localized CT image data sets.
Z-0956-2015January 21, 2015Feldkirchen
Ortho-Clinical Diagnostics
Class IITerminated
VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Catalog Number 6802445, IVD --- Ortho Clinical Diagnostics. The VITROS 4600 Chemistry System is intended for use in the in vitro q...
Reason: Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibr...
Z-0968-2015January 21, 2015Rochester, NY
St Jude Medical Cardiac Rhythm Management Division
Class IITerminated
Merlin@home RF Remote Monitoring Transmitter, Model EX1150, with software versions 6.1M or 7.0. Usage:RF Remote Transmitter
Reason: Transmitter may initiate a software reset resulting in backup operation in some implanted St. Jude Medical Radio Frequency (RF) enabled Implantable Cardioverter Assura, Unify Assur...
Z-0949-2015January 21, 2015Sylmar, CA
Ortho-Clinical Diagnostics
Class IITerminated
VITROS 5600 Integrated System, Catalog Number 6802413, IVD --- Ortho Clinical Diagnostics. For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analy...
Reason: Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibr...
Z-0969-2015January 21, 2015Rochester, NY
Ortho-Clinical Diagnostics
Class IITerminated
VITROS 3600 Immunodiagnostic System, Catalog Number 6802783, IVD --- Ortho Clinical Diagnostics. For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety o...
Reason: Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibr...
Z-0967-2015January 21, 2015Rochester, NY
Phadia US Inc
Class IITerminated
Phadia 1000 Instrument (introduced as UniCAP 1000), Article number: 12-3800-01 (All instrument software versions since launch in 2003); Multiple ImmunoCAP assays for Allergen Testing, FDA Cleared for ...
Reason: During an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a misma...
Z-0947-2015January 21, 2015Portage, MI
Brainlab AG
Class IITerminated
iPlan RT is a radiation treatment planning system that is intended for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial les...
Reason: iPlan RT Radiation Treatment Planning Software: Potentially incorrect patient positioning when using multiple localized CT image data sets.
Z-0957-2015January 21, 2015Feldkirchen
Siemens Medical Solutions USA, Inc.
Class IITerminated
The Symbia Intevo Excel is a non-diagnostic SPECT ICT system with CT support for only attenuation correction and anatomical localization. The Symbia Intevo Series are xSPECT systems. These are SPE...
Reason: Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression re...
Z-0913-2015January 14, 2015Hoffman Estates, IL
McKesson Israel Ltd.
Class IITerminated
McKesson Cardiology ECG Management It is a software application designed to import, display, store, analyze, distribute and manage information related to ECG procedures of adult and pediatric patie...
Reason: An ECG procedure time stamp is incorrect, due to a software error in the McKesson Cardiology ECG Management with software versions 13.1 and 13.1.1.
Z-0910-2015January 14, 2015Tel Aviv, N/A
Siemens Medical Solutions USA, Inc.
Class IITerminated
The Siemens Symbia S series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and or...
Reason: Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression re...
Z-0911-2015January 14, 2015Hoffman Estates, IL
Vision Rt Inc
Class IITerminated
AlignRT is a video-based three-dimensional (3D) surface imaging system which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment.
Reason: Potential use error when the external Gate Controller is turned ON after the Vision RT software has been opened.
Z-0934-2015January 14, 2015Columbia, MD
Siemens Medical Solutions USA, Inc.
Class IITerminated
The Siemens Symbia T series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and or...
Reason: Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression re...
Z-0912-2015January 14, 2015Hoffman Estates, IL
Hansen Medical Inc
Class IITerminated
Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic ...
Reason: The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolle...
Z-0896-2015January 7, 2015Mountain View, CA
Bausch & Lomb Inc
Class IITerminated
Bausch & Lomb Stellaris PC w/ Laser Upgrade BL14334 (Stellaris PC w v4 Software) with BL2272 (Laser Upgrade kit) Stellaris PC Vision Enhancement System with integrated laser module. System is use...
Reason: A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.
Z-0864-2015December 31, 2014Saint Louis, MO
Illumina Inc
Class IITerminated
Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripher...
Reason: It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot repor...
Z-0849-2015December 31, 2014San Diego, CA
Bausch & Lomb Inc
Class IITerminated
Bausch & Lomb Stellaris PC w/ Laser Upgrade BL1433 (Stellaris PC w v3 Software) with BL2272 (Laser Upgrade kit) Stellaris PC Vision Enhancement System with integrated laser module. System is used f...
Reason: A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.
Z-0863-2015December 31, 2014Saint Louis, MO
Horiba Instruments Inc
Class IITerminated
Lite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D The Lite DM (Lite Data Manager) is a software application developed by HORIBA Medical int...
Reason: HORIBA Medical is informing all HORIBA Medical Lite DM customers using software version 2.0.0.19 and lower of a software bug related to the potential for results received being app...
Z-0856-2015December 31, 2014Irvine, CA
Ge Healthcare It
Class IITerminated
Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM pr...
Reason: GE Healthcare has become aware of a potential data loss issue associated with empty path names resulting from HL-7 updates when the cache disk is almost full.
Z-0833-2015December 31, 2014Barrington, IL
Illumina Inc
Class IIITerminated
Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform.
Reason: Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrumen...
Z-0850-2015December 31, 2014San Diego, CA
Bausch & Lomb Inc
Class IITerminated
Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated Rochester, NY
Reason: A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.
Z-0862-2015December 31, 2014Saint Louis, MO
Philips Ultrasound, Inc.
Class IITerminated
Q-Station Quantification Software. Q-Station 3.0 with a2DQ and/or aCMQ Q-Apps are affected. Model: Q-Station, Catalog number: 795088, and Part Number: 989605382391. Intended to manage, view, analy...
Reason: When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ve...
Z-0818-2015December 24, 2014Bothell, WA
Carestream Health Inc
Class IITerminated
DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray Sy...
Reason: While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution M...
Z-0800-2015December 24, 2014Rochester, NY
Philips Ultrasound, Inc.
Class IITerminated
Additional devices identified in new FSN MA-FCO 83000190: QLAB versions 10.0 with a2DQ and/or aCMQ plug-ins installed with Philips Xcelera software. *******************************************...
Reason: When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ve...
Z-0815-2015December 24, 2014Bothell, WA
Philips Ultrasound, Inc.
Class IITerminated
EPIQ 5 Ultrasound System, EPIQ 5 systems with software versions 1.0.x or 1.1.x with QLAB a2DQ and/or aCMQ plug-ins are affected. Model: EPIQ 5G, EPIQ 5C, EPIQ 5W. Catalog Number: 795204 / 795205 and ...
Reason: When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ve...